Specific Process Knowledge/Characterization/Topographic measurement: Difference between revisions

From LabAdviser
BGE (talk | contribs)
Kn (talk | contribs)
Line 13: Line 13:
{| border="2" cellspacing="0" cellpadding="4" align="center"
{| border="2" cellspacing="0" cellpadding="4" align="center"
!
!
![[Specific Process Knowledge/Characterization/Profiler#Dektak_8_stylus_profiler|Dektak 8 stylus profiler]]
![[Specific Process Knowledge/Characterization/Profiler#Dektak_8_stylus_profiler|Dektak 8 stylus profiler]]
![[Specific Process Knowledge/Characterization/Profiler#Dektak_8_stylus_profiler|Dektak 8 stylus profiler]]
![[Specific Process Knowledge/Characterization/Profiler#Optical_Profiler_(Sensofar)|Optical Profiler (Sensofar)]]
![[Specific Process Knowledge/Characterization/Profiler#Optical_Profiler_(Sensofar)|Optical Profiler (Sensofar)]]
Line 18: Line 19:
|- valign="top"
|- valign="top"
|'''General description'''
|'''General description'''
|Profiler for measuring micro structures. Can do wafer mapping and stress measurements.
|Profiler for measuring micro structures. Can do wafer mapping and stress measurements.
|Profiler for measuring micro structures. Can do wafer mapping and stress measurements.
|3D Profiler for measuring micro structures. Can do wafer mapping.
|3D Profiler for measuring micro structures. Can do wafer mapping.
Line 24: Line 26:
|'''Substrate size'''
|'''Substrate size'''
|up to 8"
|up to 8"
|up to 6"
|Up to more than 6"
|Up to more than 6"
|6" or less
|6" or less
|-valign="top"
|-valign="top"
|'''Max. scan range xy'''
|'''Max. scan range xy'''
|Line scan x: 50µm to 200mm  
|Line scan x: 50µm to 200mm
|Line scan x: 50µm to 200mm 
|Depending on the objective:
|Depending on the objective:
*One view: 127µmX95µm to 1270µmX955µm
*One view: 127µmX95µm to 1270µmX955µm
Line 35: Line 39:
|-valign="top"
|-valign="top"
|'''Max. scan range z'''
|'''Max. scan range z'''
|50Å to 262µm
|50Å to 1mm
|50Å to 1mm
|Depending on the objective and Z resolution:
|Depending on the objective and Z resolution:
*94.4µm ->9984µm
*94.4µm ->9984µm
Line 42: Line 47:
|'''Resolution xy'''
|'''Resolution xy'''
|down to 0.067 µm
|down to 0.067 µm
|down to xxx µm
|Depending on the objective:
|Depending on the objective:
*0.5µm -> 5µm
*0.5µm -> 5µm
Line 47: Line 53:
|-valign="top"
|-valign="top"
|'''Resolution z'''
|'''Resolution z'''
|1Å, 10Å or 20Å
|1Å, 10Å or 20Å
|1Å, 10Å or 20Å
|Depending on measuring methode:
|Depending on measuring methode:
Line 55: Line 62:
|- valign="top"
|- valign="top"
|'''Max. scan depth [µm] (as a function of trench width W''')
|'''Max. scan depth [µm] (as a function of trench width W''')
|1.2*(W[µm]-5µm)
|1.2*(W[µm]-5µm)
|1.2*(W[µm]-5µm)
|Depending on material and trench width:
|Depending on material and trench width:
Line 61: Line 69:
|-valign="top"
|-valign="top"
|'''Tip radius'''
|'''Tip radius'''
|5 µm 45<sup>o</sup> cone
|5 µm 45<sup>o</sup> cone
|5 µm 45<sup>o</sup> cone
|No tip - using light
|No tip - using light
Line 68: Line 77:
|-valign="top"
|-valign="top"
|'''Stress measurement'''
|'''Stress measurement'''
|Can be done
|Can be done
|Can be done
|No stress calculation capability
|No stress calculation capability
Line 73: Line 83:
|-valign="top"
|-valign="top"
|'''Surface roughness'''
|'''Surface roughness'''
|Can be done on a line scan
|Can be done on a line scan
|Can be done on a line scan
|Can be done on a line or an area
|Can be done on a line or an area

Revision as of 15:21, 3 October 2012

Topographic measurements are measurements were you can measure hight differences on your substrate. If you measure in many spots of the substrate you can get a topographic image of the substrate.

AT DANCHIP we have three systems for topographic measurement:

  • Dektak - Profiler for measuring micro structures
  • Optical Profiler (Sensofar) - 3D Profiler for measuring micro structures
  • Nanoman - AFM for measuring nano structures

High Aspect ratio structures
The fact that the tips of the Dektak profilers and the AFM are shaped like a cone with some tip angle causes a problem when characterizing high aspect ratio structures. For instance, if a 20 m wide trench is etched deeper than approximately 18 m, the tip of the Dektak will not be able to reach the bottom. The optical profiler uses a light beam that is focused through an objective. Therefore it is able to measure higher aspect ratios. The aspect ratio is limited by the posibility for the light to reach the bottom and get back to the detector. On some samples we have been able to measure aspect ratios above 1:10. Otherwise the solution is to cleave the sample along a line that is perpendicular to the trench and then inspect the profile in a scanning electron microscope or a microscope (for large structures).


Comparison of the two profilers and the AFM

Dektak 8 stylus profiler Dektak 8 stylus profiler Optical Profiler (Sensofar) Nanoman
General description Profiler for measuring micro structures. Can do wafer mapping and stress measurements. Profiler for measuring micro structures. Can do wafer mapping and stress measurements. 3D Profiler for measuring micro structures. Can do wafer mapping. AFM for measuring nanostructures and surface roughness
Substrate size up to 8" up to 6" Up to more than 6" 6" or less
Max. scan range xy Line scan x: 50µm to 200mm Line scan x: 50µm to 200mm Depending on the objective:
  • One view: 127µmX95µm to 1270µmX955µm
  • Stitching: In principel a hole 6" wafer (time consuming)
90 µm square
Max. scan range z 50Å to 1mm 50Å to 1mm Depending on the objective and Z resolution:
  • 94.4µm ->9984µm
1 µm (can go up to 5 µm under special settings)
Resolution xy down to 0.067 µm down to xxx µm Depending on the objective:
  • 0.5µm -> 5µm
Depending on scan size and number of samples per line and number of lines - accuracy better than 2%
Resolution z 1Å, 10Å or 20Å 1Å, 10Å or 20Å Depending on measuring methode:
  • PSI down to 0.01 nm
  • VSI down to 1 nm
  • Confocal (depending on objective): 1nm -> 50nm
<1Å - accuracy better than 2%
Max. scan depth [µm] (as a function of trench width W) 1.2*(W[µm]-5µm) 1.2*(W[µm]-5µm) Depending on material and trench width:
  • Somewhere between 1:1 and 1:12
~1:1 with standard cantilever.
Tip radius 5 µm 45o cone 5 µm 45o cone No tip - using light
  • Blue monochromatic LED: 460nm
  • White broadband LED: 550nm
<12 nm on standard cantilever
Stress measurement Can be done Can be done No stress calculation capability Cannot be done
Surface roughness Can be done on a line scan Can be done on a line scan Can be done on a line or an area Can be done on a selected surface area