Line 109: Line 109:
*TSA microscope travel range: X ±25 mm; Y +20/-75 mm (towards the flat)
*TSA microscope travel range: X ±25 mm; Y +20/-75 mm (towards the flat)


'''BackSide Alignment:'''
 
'''Back Side Alignment:'''
*Minimum distance between BSA microscope objectives: 15 mm
*Minimum distance between BSA microscope objectives: 15 mm
*Maximum distance between BSA microscope objectives: 100 mm
*Maximum distance between BSA microscope objectives: 100 mm
Line 120: Line 121:


'''Microscope field of view (W x H, splitfield):'''
'''Microscope field of view (W x H, splitfield):'''
*TSA 5X
*TSA 5X:
**Oculars: 1.3mm x 2.6mm (Ø2.6mm full field)
**Oculars: 1.3 mm x 2.6 mm (Ø2.6 mm full field)
**Camera: 350µm x 500µm (700µm x 500µm full field)
**Camera: 350 µm x 500 µm (700 µm x 500 µm full field)
*TSA 10X
*TSA 10X:
**Oculars: 0.6mm x 1.3mm (Ø1.3mm full field)
**Oculars: 0.6 mm x 1.3 mm (Ø1.3 mm full field)
**Camera: 150µm x 250µm (350µm x 250µm full field)  
**Camera: 150 µm x 250 µm (350 µm x 250 µm full field)  
*TSA special
*TSA special:
**Oculars: 0.55mm x 1.1mm (Ø1.1mm full field)
**Oculars: 0.55 mm x 1.1 mm (Ø1.1 mm full field)
**Camera: 150µm x 200µm (300µm x 200µm full field)
**Camera: 150 µm x 200 µm (300 µm x 200 µm full field)
*BSA camera
*BSA camera:
**Low: 1.5mm x 2mm (3mm x 2mm full field)
**Low: 1.5 mm x 2 mm (3 mm x 2 mm full field)
**High: 450µm x 650µm (950µm x 650µm full field)
**High: 450 µm x 650 µm (950 µm x 650 µm full field)


==Quality Control (QC)==
==Quality Control (QC)==