Jump to content

Specific Process Knowledge/Lithography/EBeamLithography/HSQ Dose Test: Difference between revisions

From LabAdviser
Rawta (talk | contribs)
Rawta (talk | contribs)
Line 12: Line 12:
Below is overview optical images of the developed pattern:
Below is overview optical images of the developed pattern:
[[File:Dose test optical.png|frameless|625x625px|dose test optical images|left]]
[[File:Dose test optical.png|frameless|625x625px|dose test optical images|left]]


SEM images were recorded and analyzed from the structure shown below:
SEM images were recorded and analyzed from the structure shown below:
[[File:design.png|frameless|625x625px|dose test optical images|left]]
[[File:design.png|frameless|383x383px|dose test optical images|left]]
 
 
 
 
 
 


The measured and designed critical dimension are plotted vs the doses below:
The measured and designed critical dimension are plotted vs the doses below:
[[File:CD vs dose.png|frameless|625x625px|dose test optical images|left]]
[[File:CD vs dose.png|frameless|446x446px|dose test optical images|left]]


The line width roughness (LWR) is plotted vs CD below:  
The line width roughness (LWR) is plotted vs CD below:  
[[File:LWR vs CD.png|frameless|625x625px|dose test optical images|left]]
[[File:LWR vs CD.png|frameless|462x462px|dose test optical images|left]]


.
.

Revision as of 14:14, 9 September 2025

Resist info

  • we used a 10 % HSQ resist for a dose test. The resist info sheet could be found her.
  • spinning: in labspin 3000rpm for 40s with 2s ramp (1500 acceleration) then bake for 2 min at 165 C
  • Exposure: At 12nA ap6 (spot size 10 nm) in JEOL 9500
    • Dose test with structures (3x3)
    • Develop for 90s using manual TMAH developer
    • Optical and SEM images and analysis  

Results

Below is overview optical images of the developed pattern:

dose test optical images
dose test optical images










SEM images were recorded and analyzed from the structure shown below:

dose test optical images
dose test optical images




The measured and designed critical dimension are plotted vs the doses below:

dose test optical images
dose test optical images

The line width roughness (LWR) is plotted vs CD below:

dose test optical images
dose test optical images

.