Specific Process Knowledge/Lithography/Strip: Difference between revisions
Appearance
| Line 298: | Line 298: | ||
==Process power for plasma asher 4 & 5== | ==Process power for plasma asher 4 & 5== | ||
[[File: | [[File:PA_power_v2.png|320px|thumb|Ashing rate as function of microwave power.|right]] | ||
The ashing rate is related to the power used during processing. Higher power gives higher ashing rate. | The ashing rate is related to the power used during processing. Higher power gives higher ashing rate. | ||
Test using a single 100 mm wafer in the center of the process chamber shows that a power of 1000 W gives the highest ashing rate. | Test using a single 100 mm wafer in the center of the process chamber shows that a power of 1000 W gives the highest ashing rate. | ||