*PID tables - This table contains the necessary coefficients that regulate the power output, in a smooth and control way, to keep the temperature at the desired setpoint while processing. | *PID tables - This table contains the necessary coefficients that regulate the power output, in a smooth and control way, to keep the temperature at the desired setpoint while processing. | ||
*Pyro Cal Tables - The optical pyrometer needs to be calibrated using a thermocouple as a reference, as optical pyrometry is | *Pyro Cal Tables - The optical pyrometer needs to be calibrated using a thermocouple as a reference, as optical pyrometry is a non-contact method for measuring the temperature inside the chamber. As a result of such calibration process, a table is created. The pyrometer calibration table correlates temperatures recorded by the reference thermocouple with the electrical signals produced by the pyrometer. Essentially, this table functions as a 'translator,' indicating to the user the temperature corresponding to a specific output from the pyrometer. | ||
==Additional information== | ==Additional information== | ||
The '''user manual''', user '''APV''', '''technical documents and contact information''' can be found in LabManager: '''[http://www.labmanager.dtu.dk/function.php?module=Machine&view=view&mach=459 RTP2 Jipelec]''' | The '''user manual''', user '''APV''', '''technical documents and contact information''' can be found in LabManager: '''[http://www.labmanager.dtu.dk/function.php?module=Machine&view=view&mach=459 RTP2 Jipelec]''' | ||