Jump to content

Specific Process Knowledge/Lithography/EBeamLithography/JEOL 9500 User Guide: Difference between revisions

Thope (talk | contribs)
No edit summary
Thope (talk | contribs)
No edit summary
Line 1: Line 1:
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Lithography/EBeamLithography/JEOL 9500 User Guide click here]'''
Content and illustration by Thomas Pedersen, DTU Nanolab unless otherwise noted.
[[File:TPE02803.jpg|right|400px]]
[[File:TPE02803.jpg|right|400px]]


Line 72: Line 76:
|-  
|-  
| colspan="2" style="text-align: center;|
| colspan="2" style="text-align: center;|
Operation screen of the automatic cassette transfer system / auto stocker. Photo: Thomas Pedersen.
Operation screen of the automatic cassette transfer system / auto stocker.
|}
|}


Line 95: Line 99:
|-  
|-  
| colspan="2" style="text-align: center;|
| colspan="2" style="text-align: center;|
Samples can be loaded into appropriate cassettes on the cassette preparation table. Photo: Thomas Pedersen.
Samples can be loaded into appropriate cassettes on the cassette preparation table.
|}
|}


Line 109: Line 113:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
Nanolab staff member loading the 4" wafer cassette to the automatic cassette transfer system. Photo: Thomas Pedersen.
Nanolab staff member loading the 4" wafer cassette to the automatic cassette transfer system.
|}
|}


Line 190: Line 194:
|-  
|-  
| colspan="2" style="text-align:center;|
| colspan="2" style="text-align:center;|
'''Calibration''' and '''RESTOR''' windows. Image: Thomas Pedersen.
'''Calibration''' and '''RESTOR''' windows.
|}
|}


Line 205: Line 209:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
Result display of current measurement. Image: Thomas Pedersen.
Result display of current measurement.
|}
|}


Line 221: Line 225:
|-  
|-  
| colspan="2" style="text-align:center;|
| colspan="2" style="text-align:center;|
Correct result of '''INITAE''' (left) and '''INITBE''' (right). Image: Thomas Pedersen.
Correct result of '''INITAE''' (left) and '''INITBE''' (right).
|}
|}


Line 240: Line 244:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
Load the '''daily''' batch command and execute it. Image: Thomas Pedersen.
Load the '''daily''' batch command and execute it.
|}
|}


Line 267: Line 271:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
Comparison of two drift measurements. Image: Thomas Pedersen.
Comparison of two drift measurements.
|}
|}


Line 292: Line 296:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
Parameter window of the '''HEIMAP''' subprogram. Image: Thomas Pedersen.
Parameter window of the '''HEIMAP''' subprogram.
|}
|}


Line 312: Line 316:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
Acquire the calibration data and then apply the data to save. Image: Thomas Pedersen.
Acquire the calibration data and then apply the data to save.
|}
|}


Line 335: Line 339:
|-  
|-  
| colspan="1" style="text-align:center;|
| colspan="1" style="text-align:center;|
'''Expose''' window with .mgn file loaded for exposure. Notice that the '''Progress''' part of the window still shows the previous exposure information. This field will not update until exposure is started. Image: Thomas Pedersen.