Jump to content

Specific Process Knowledge/Etch/DRIE-Pegasus/Pegasus-2: Difference between revisions

Jmli (talk | contribs)
Jmli (talk | contribs)
Line 80: Line 80:
}}
}}
{{Template:Peg2configcontent1
{{Template:Peg2configcontent1
|ItemName= Electrostatic clamping, chip bonding with Galden oil and helium backside cooling are no longer being used.
|ItemName= Electrostatic clamping and helium backside cooling are again mandatory.
|ItemConfiguration= The processes run in Pegasus 2 operate at such low powers that they do not require very efficient cooling. It has therefore been decided that the following options/procedures
|ItemConfiguration= The processes run in Pegasus 2 operate at such low powers that they do not require very efficient cooling. It has therefore been decided that the following options/procedures
* Electrostatic clamping of carriers/substrates
* Electrostatic clamping of carriers/substrates
* Helium backside cooling
* The Helium backside cooling is kept at 5 Torr instead of the usual 10 Torr on the other Pegasus tools.
* Using bonding procedures such as droplets of Galden oil to increase thermal contact between chips and carrier
* Using bonding procedures such as droplets of Galden oil to increase thermal contact between chips and carrier
will no longer be required.
will no longer be required.