Specific Process Knowledge/Etch/DRIE-Pegasus/Pegasus-2/ORE with Al2O3 mask: Difference between revisions
Appearance
| Line 13: | Line 13: | ||
<br> | <br> | ||
Moreover, during the E-step, the MFC (mass flow controller) presented a delay to read the pressure when compared with the power, creating unwanted bias. To fight it, the E-step was divided into E<sub>1</sub> and E<sub>2</sub>. The E<sub>1</sub> of only 2 s is enough to stabilize the pressure, and E<sub>2</sub> the profile is etch correctly, without the unwanted bias. | Moreover, during the E-step, the MFC (mass flow controller) presented a delay to read the pressure when compared with the power, creating unwanted bias. To fight it, the E-step was divided into E<sub>1</sub> and E<sub>2</sub>. The E<sub>1</sub> of only 2 s is enough to stabilize the pressure, and E<sub>2</sub> the profile is etch correctly, without the unwanted bias. | ||
[[File:graf bias & recipes.png| | [[File:graf bias & recipes.png|600px|center|thumb|'''''CORE process graphics and recipes: a) unwanted DC bias during the E-step; b) E-step separated into two steps, E<sub>1</sub> decreased the bias.''''']] | ||
=== Pillars === | === Pillars === | ||
| Line 41: | Line 41: | ||
|+ 1μm holes recipe from March 2022 - #288 = 120 min | |+ 1μm holes recipe from March 2022 - #288 = 120 min | ||
|- | |- | ||
|-style="background: | |-style="background:LightGray; color:Black" | ||
! !! Time (s) !! Pressure !! O<sub>2</sub> flow (SCCM) !! SF<sub>6</sub> flow (SCCM) !! Platen power (W) !! Coil power (W) | ! !! Time (s) !! Pressure !! O<sub>2</sub> flow (SCCM) !! SF<sub>6</sub> flow (SCCM) !! Platen power (W) !! Coil power (W) | ||
|- | |- | ||
| Line 60: | Line 60: | ||
|+ 200nm nanoholes recipe from March 2022 | |+ 200nm nanoholes recipe from March 2022 | ||
|- | |- | ||
|-style="background: | |-style="background:LightGray; color:Black" | ||
! !! Time (s) !! Pressure (valve control) !! O<sub>2</sub> flow (SCCM) !! SF<sub>6</sub> flow (SCCM) !! Platen power (W) | ! !! Time (s) !! Pressure (valve control) !! O<sub>2</sub> flow (SCCM) !! SF<sub>6</sub> flow (SCCM) !! Platen power (W) | ||
|- | |- | ||