Jump to content

Specific Process Knowledge/Characterization/Profiler: Difference between revisions

Reet (talk | contribs)
Reet (talk | contribs)
Line 152: Line 152:
|style="background:LightGrey; color:black"|Resolution Z
|style="background:LightGrey; color:black"|Resolution Z
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
0.01 Å, 0.08 Å, or 0.6 Å according to the manufacturer for ranges 13 µm, 131 µm, and 1 mm.*
0.01 Å, 0.08 Å, or 0.6 Å according to the manufacturer for ranges 13 µm, 131 µm, and 1 mm.
|-
|-
|style="background:LightGrey; color:black"|Height accuracy z (95 % confidence)
|style="background:LightGrey; color:black"|Height accuracy z (95 % confidence)
Line 179: Line 179:
|}
|}


*The resolution for the smallest range is theoretical, as this is below the noise threshold at least in our lab.
†The resolution for the smallest range is theoretical, as this is below the noise threshold at least in our lab.


===Height measurement accuracy for the Tencor P17 Stylus Profiler===
===Height measurement accuracy for the Tencor P17 Stylus Profiler===