LabAdviser/314/Microscopy 314-307/TEM/ETEM/dose-calibration: Difference between revisions

From LabAdviser
< LabAdviser‎ | 314‎ | Microscopy 314-307‎ | TEM‎ | ETEM
Jenk (talk | contribs)
mNo edit summary
Jenk (talk | contribs)
(No difference)

Revision as of 07:59, 28 May 2020

Set dose calibration for OneView camera in Digital Micrograph


Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for the OneView camera needs to be adjusted. This is done by running an installed script:

1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU"

Run script


2. Type in the high tension the microscope is running in the pop-up window

Select high tension


3. Don't change the conversion factor; just click on "OK"

Don't change conversion factor



Feedback to this page: click here