== For the next user: Changing the beam kV and settings for the alignment and FEG register == | |||
# Go to Set up → High Tension → Decrease/Increase the HT in steps from 300 kV/120 kV to the needed setting <br/><br/> | |||
# Go to Alignments tab → Alignments → Flapout → select the required setting (e.g. Main_120kV) and then press Ctrl+A for selecting all and then move them from “Available” to “Selected” by clicking on the Apply button. <br/><br/> | |||
# Go to Gun → FEG Register → Load the FEG register for the electron energy of interest to the next user <br/><br/> | |||
# Go to Setup tab → Vacuum → Flapout → Click on the cryo cycle button and it will turn yellow <br/><br/> | |||
# Stop your session in Lab Manager <br/><br/> | |||
# Log out from lab manager and the support PC. | |||