== For the next user: Changing the beam kV and settings for the alignment and FEG register ==
# Go to Set up &rarr; High Tension &rarr; Decrease/Increase the HT in steps from 300 kV/120 kV to the needed setting <br/><br/>
# Go to Alignments tab &rarr; Alignments &rarr; Flapout &rarr; select the required setting (e.g. Main_120kV) and then press Ctrl+A for selecting all and then move them from “Available” to “Selected” by clicking on the Apply button. <br/><br/>
# Go to Gun &rarr; FEG Register &rarr; Load the FEG register for the electron energy of interest to the next user <br/><br/>
# Go to Setup tab &rarr; Vacuum &rarr; Flapout &rarr; Click on the cryo cycle button and it will turn yellow <br/><br/>
# Stop your session in Lab Manager <br/><br/>
# Log out from lab manager and the support PC.