LabAdviser/314/Microscopy 314-307/TEM/ETEM/dose-calibration: Difference between revisions

From LabAdviser
< LabAdviser‎ | 314‎ | Microscopy 314-307‎ | TEM‎ | ETEM
Jenk (talk | contribs)
Created page with "'''Set dose calibration for OneView camera in Digital Micrograph''' Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for t..."
 
Jenk (talk | contribs)
No edit summary
Line 7: Line 7:
1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU" <br />
1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU" <br />
[[File:Oneview_1.png|600px|center|thumb|Run script]]
[[File:Oneview_1.png|600px|center|thumb|Run script]]
<br />
<br clear="all" />


2. Type in the high tension the microscope is running in the pop-up window<br />
2. Type in the high tension the microscope is running in the pop-up window<br />
[[File:Oneview_2.png|300px|center|thumb|Select high tension]]
[[File:Oneview_2.png|300px|center|thumb|Select high tension]]
<br />
<br clear="all" />


2. Don't change the conversion factor; just click on "OK"<br />
2. Don't change the conversion factor; just click on "OK"<br />
[[File:Oneview_3.png|300px|center|thumb|Don't change conversion factor]]
[[File:Oneview_3.png|300px|center|thumb|Don't change conversion factor]]
<br />
<br clear="all" />




'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/314/ETEM/dose-calibration click here]'''
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/314/ETEM/dose-calibration click here]'''

Revision as of 13:43, 26 March 2020

Set dose calibration for OneView camera in Digital Micrograph


Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for the OneView camera needs to be adjusted. This is done by running an installed script:

1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU"

Run script


2. Type in the high tension the microscope is running in the pop-up window

Select high tension


2. Don't change the conversion factor; just click on "OK"

Don't change conversion factor



Feedback to this page: click here