Created page with "'''Set dose calibration for OneView camera in Digital Micrograph''' Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for t..."
 
Jenk (talk | contribs)
No edit summary
Line 7: Line 7:
1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU" <br />
1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU" <br />
[[File:Oneview_1.png|600px|center|thumb|Run script]]
[[File:Oneview_1.png|600px|center|thumb|Run script]]
<br />
<br clear="all" />


2. Type in the high tension the microscope is running in the pop-up window<br />
2. Type in the high tension the microscope is running in the pop-up window<br />
[[File:Oneview_2.png|300px|center|thumb|Select high tension]]
[[File:Oneview_2.png|300px|center|thumb|Select high tension]]
<br />
<br clear="all" />


2. Don't change the conversion factor; just click on "OK"<br />
2. Don't change the conversion factor; just click on "OK"<br />
[[File:Oneview_3.png|300px|center|thumb|Don't change conversion factor]]
[[File:Oneview_3.png|300px|center|thumb|Don't change conversion factor]]
<br />
<br clear="all" />




'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/314/ETEM/dose-calibration click here]'''