==Filmtek 4000== | |||
[[image:Filmtek.JPG|275x275px|right|thumb|FilmTek 4000: positioned in cleanroom C-1]] | |||
FilmTek 4000 is a computerized film thickness measurement and material characterization system. This system combines fiber-optic spectrophotometry with advanced material modeling software to provide an affordable and reliable tool for the simultaneous measurement of film thickness, index of refraction, and extinction coefficient | |||
Normal incidence and polarized 70 degree reflection data is collected and used to calculate thickness and index of refraction of the measured film using SCI’s patented Differential Power Spectral Density (DPSD) technique. Absolute reflection data is obtained by comparing sample data to the measured reflection of a known reference sample, typically a silicon wafer with a thermally grown thin oxide (~63 nm) film. | |||
The spectrophotometer scans the sample over a predefined range of wavelengths. The software generates a reflection spectrum based on a previously stored reference scan, and then performs a regression on the unknown parameters to fit the simulated reflection and power spectral density to the observed values. The resulting thin film parameters along with the measured and modeled spectra are then displayed for the user to examine. | |||
'''The user manual, quality control procedure and results, technical information and contact information can be found in LabManager:''' | |||
[http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=169 FilmTek in LabManager] | |||
<br clear="all" /> | |||
===An overview of the performance of the FilmTek=== | |||
{| border="2" cellspacing="0" cellpadding="10" | |||
|- | |||
!style="background:silver; color:black;" align="left"|Purpose | |||
|style="background:LightGrey; color:black"|Film thickness measurements and optical characterization of optically transparent thin films||style="background:WhiteSmoke; color:black"| | |||
*Measurement of (multi layer) film thickness (only one unknown layer) | |||
*Optical constants | |||
*Surface roughness | |||
|- | |||
!style="background:silver; color:black" align="left"|Performance | |||
|style="background:LightGrey; color:black"|Thin film materials that can be measured||style="background:WhiteSmoke; color:black"| | |||
Any film that is transparent to the light in the given wavelength range | |||
ex: | |||
*Silicon Oxide | |||
*Silicon nitride | |||
*PolySilicon | |||
*Photoresists | |||
*SU8 | |||
*Other polymers | |||
*Very thin layers of metals (<20 nm) | |||
*and many more | |||
|- | |||
|style="background:silver; color:black"| | |||
|style="background:LightGrey; color:black"|Film thickness range | |||
|style="background:WhiteSmoke; color:black"| | |||
*<100 Å to 250 µm (depending of the material) | |||
|- | |||
!style="background:silver; color:black" align="left"|Process parameter range | |||
|style="background:LightGrey; color:black"|Wavelength range | |||
|style="background:WhiteSmoke; color:black"| | |||
*400-1000 nm | |||
|- | |||
!style="background:silver; color:black" align="left"|Substrates | |||
|style="background:LightGrey; color:black"|Batch size | |||
|style="background:WhiteSmoke; color:black"| | |||
*One sample at a time - all sample larger than 5x5 mm<sup>2</sup>sizes up to 6" | |||
|- | |||
|style="background:silver; color:black"| | |||
| style="background:LightGrey; color:black"|Substrate material allowed | |||
|style="background:WhiteSmoke; color:black"| | |||
*In principle all materials | |||
*Only pure silicon, silicon oxides, silicon nitrides and quartz may be in direct contact with the surface of the stage. If you have metals, III-V materials or polymers on the back side of the substrate the please mount your sample on a silicon carrier wafer. | |||
|- | |||
|} | |||
<br clear="all" /> | |||
<!-- | |||
==Prism Coupler== | |||
[[image:Prism_coupler.jpg|300x300px|right|thumb|Prism coupler: (old image), now positioned cleanroom D-3]] | |||
The Prism coupler is a Metricon model 2010 Prism Coupler. It is an old system from 1991 but the system with a few modifications is still beening manufactured today. It is still a very easy and relaible method of determining thickness and refractive index. | |||
The prism coupler uses an optical waveguide technique to determine the thickness and refractive index of a thin film. To learn more about the theory of measurement please see the homepage of [http://www.metricon.com Metricon] | |||
It is especially good for making fast accurate measurements of dielectric thin films with n<2.02 and a thickness range of 1 µm to about 15 µm. | |||
'''The user manual, quality control procedure and results, technical information and contact information can be found in LabManager:''' | |||
[http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=12 Prism Coupler in LabManager] | |||
<br clear="all" /> | |||
===An overview of the performance of the Prism Coupler=== | |||
{| border="2" cellspacing="0" cellpadding="10" | |||
|- | |||
!style="background:silver; color:black;" align="left"|Purpose | |||
|style="background:LightGrey; color:black"|Film thickness measurements and optical characterization of optically transparent thin films||style="background:WhiteSmoke; color:black"| | |||
*Measurement of film thickness | |||
*Optical constants | |||
|- | |||
!style="background:silver; color:black" align="left"|Performance | |||
|style="background:LightGrey; color:black"|Thin film materials that can be measured||style="background:WhiteSmoke; color:black"| | |||
Film with a refractive index of less than 2.02 and that are transparent to the light in the given wavelength range | |||
ex: | |||
*Silicon Oxide | |||
*Silicon nitride | |||
*polymers | |||
|- | |||
|style="background:silver; color:black"| | |||
|style="background:LightGrey; color:black"|Film thickness range | |||
|style="background:WhiteSmoke; color:black"| | |||
*~1µm to 15 µm | |||
|- | |||
|style="background:silver; color:black"| | |||
|style="background:LightGrey; color:black"|Film thickness accuracy | |||
|style="background:WhiteSmoke; color:black"| | |||
*±(0.5%+50Å) | |||
|- | |||
|style="background:silver; color:black"| | |||
|style="background:LightGrey; color:black"|Index accuracy | |||
|style="background:WhiteSmoke; color:black"| | |||
*±0.001 | |||
|- | |||
!style="background:silver; color:black" align="left"|Process parameter range | |||
|style="background:LightGrey; color:black"|Wavelength range | |||
|style="background:WhiteSmoke; color:black"| | |||
Can operate at two different wavelength: | |||
*633 nm | |||
*1550 nm (not working) | |||
|- | |||
!style="background:silver; color:black" align="left"|Substrates | |||
|style="background:LightGrey; color:black"|Batch size | |||
|style="background:WhiteSmoke; color:black"| | |||
*One sample at a time - all sample larger than 5x5 mm<sup>2</sup>sizes up to 6" | |||
|- | |||
|style="background:silver; color:black"| | |||
| style="background:LightGrey; color:black"|Substrate material allowed | |||
|style="background:WhiteSmoke; color:black"| | |||
*In principle all materials | |||
|- | |||
|} | |||
--> | |||
==Filmtek 4000== | ==Filmtek 4000== |
Revision as of 11:10, 28 January 2020
Feedback to this page: click here
Comparison of the two methods
For comparison of the three methods, see here: Film thickness and optical constants of optical transparent films
Ellipsometer VASE and Ellipsometer M-2000V
The ellipsometer VASE is actually a M2000XI-210 ellipsometer from J.A. Woollam Co., Inc.. We call it VASE (Variable Angle Spectroscopic Ellipsometry) because it can do spectroscopic ellipsometry at variable angles.
The ellipsometer M2000V is an in-situ ellipsometer from J.A. Wollam Co., Inc. This dedicated for being used on the Sputter System Lesker instrument or the ALD2 for in-situ measurements. When it is not on any system it is positioned next to the ALD2.
Ellipsometry is a very sensitive characterization technique which can be used to determine thin film layer thicknesses and/or optical constants. It sends in polarized light on the surface at different angles and measures the change in polarization state of the reflected light.
It is especially good for thin films in the thickness range of a few nanometers to a few microns. The layer should be transparent to light in the usable wavelength range for the thickness to be determined. Large roughness (>40nm) and features within the field of view can cause the measurement to fail.
Ellipsometry is an indirect measurement so a model has to be fit to the data in order to obtain the film thickness and optical constants. To learn more about ellipsometry you can take a look at the tutorial provided by the J. A. Woollam Co.
Access to use the CompleteEASE software can be found using Remote Desktop connection to: 10.51.40.99 (or DCHP-D0037.win.dtu.dk)
The user manuals, quality control procedure and results and contact information can be found in LabManager:
LabManager page for Ellipsometer VASE
LabManager page for Ellipsometer M-2000V
More details
A rough overview of the performance of the Ellipsometers
Equipment | Ellipsometer VASE | Ellipsometer M-2000V | |
---|---|---|---|
Purpose |
|
| |
Performance | Films that can be examined |
Any film that is transparent to the light in the given wavelength range e.g.:
|
Any film that is transparent to the light in the given wavelength range e.g.:
|
Film thickness range |
|
| |
Process parameter range | Wavelength range |
|
|
Incident angle range |
|
| |
Beam size |
|
| |
Mapping facility |
|
| |
Accessories | Focusing optics |
|
|
Transmission stage |
|
| |
Substrates | Batch size |
|
|
Allowed materials |
|
|
Filmtek 4000
FilmTek 4000 is a computerized film thickness measurement and material characterization system. This system combines fiber-optic spectrophotometry with advanced material modeling software to provide an affordable and reliable tool for the simultaneous measurement of film thickness, index of refraction, and extinction coefficient
Normal incidence and polarized 70 degree reflection data is collected and used to calculate thickness and index of refraction of the measured film using SCI’s patented Differential Power Spectral Density (DPSD) technique. Absolute reflection data is obtained by comparing sample data to the measured reflection of a known reference sample, typically a silicon wafer with a thermally grown thin oxide (~63 nm) film.
The spectrophotometer scans the sample over a predefined range of wavelengths. The software generates a reflection spectrum based on a previously stored reference scan, and then performs a regression on the unknown parameters to fit the simulated reflection and power spectral density to the observed values. The resulting thin film parameters along with the measured and modeled spectra are then displayed for the user to examine.
The user manual, quality control procedure and results, technical information and contact information can be found in LabManager:
An overview of the performance of the FilmTek
Purpose | Film thickness measurements and optical characterization of optically transparent thin films |
|
---|---|---|
Performance | Thin film materials that can be measured |
Any film that is transparent to the light in the given wavelength range ex:
|
Film thickness range |
| |
Process parameter range | Wavelength range |
|
Substrates | Batch size |
|
Substrate material allowed |
|
Filmtek 4000
FilmTek 4000 is a computerized film thickness measurement and material characterization system. This system combines fiber-optic spectrophotometry with advanced material modeling software to provide an affordable and reliable tool for the simultaneous measurement of film thickness, index of refraction, and extinction coefficient
Normal incidence and polarized 70 degree reflection data is collected and used to calculate thickness and index of refraction of the measured film using SCI’s patented Differential Power Spectral Density (DPSD) technique. Absolute reflection data is obtained by comparing sample data to the measured reflection of a known reference sample, typically a silicon wafer with a thermally grown thin oxide (~63 nm) film.
The spectrophotometer scans the sample over a predefined range of wavelengths. The software generates a reflection spectrum based on a previously stored reference scan, and then performs a regression on the unknown parameters to fit the simulated reflection and power spectral density to the observed values. The resulting thin film parameters along with the measured and modeled spectra are then displayed for the user to examine.
The user manual, quality control procedure and results, technical information and contact information can be found in LabManager:
An overview of the performance of the FilmTek
Purpose | Film thickness measurements and optical characterization of optically transparent thin films |
|
---|---|---|
Performance | Thin film materials that can be measured |
Any film that is transparent to the light in the given wavelength range ex:
|
Film thickness range |
| |
Process parameter range | Wavelength range |
|
Substrates | Batch size |
|
Substrate material allowed |
|