Specific Process Knowledge/Bonding: Difference between revisions
No edit summary |
|||
Line 1: | Line 1: | ||
'''Feedback to this page''': '''[mailto:labadviser@ | '''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Bonding click here]''' | ||
Revision as of 12:35, 25 November 2019
Feedback to this page: click here
Choose equipment
Choose bonding methods in Wafer Bonder 2
Comparing the three bonding methods in the wafer bonder 2
Eutectic bonding | Fusion bonding | Anodic bonding | |
---|---|---|---|
General description | For bonding two substrates by use of an interphase that makes an eutecticum. | For bonding two identical materials. | For bonding Si and Glass. |
Bonding temperature | Depending on the eutecticum 310°C to 400°C. | Depending on defects 50°C to 400°C. | Depending on the voltage 300°C to 500°C Standard is 400°C. |
Annealing temperature | No annealing | 1000°C-1100°C in the anneal bond furnace (C3). | No annealing |
Materials possible to bond | Bonding of substrates is done by use of the eutectica Au/Si, Au/Sn and Au/Sn/Ni | Si/Si, SiO2/SiO2 | Si/Pyrex (glass) |
Substrate size | Up to 4" | Up to 4" | Up to 4" |
Cleaning | Cleaning by N2. | Wet chemical cleaning, IMEC. | Cleaning by N2. |
Backside alignment | Double side polished wafers. | Double side polished wafers. | Not relevant. |