In the profile below there is a few problems.
In the profile below there is one problem.


[[File:ALD-Sandwich-29a.jpg|700px]]
[[File:ALD-Sandwich-29a.jpg|700px]]


As seen above:
As seen above there is only a very small amount of Ti2p. Why is this? Let us zoom in onto the Ti2p peak and see how the fitting goes.
* The O1s content is some 80-90% until the layer containing Al2p is reached.
 
* The Hf4f is not there.
 
* There is only a very small amount of Ti2p.
[[File:ALD sandwich Ti survey.gif|700px]]

Revision as of 14:46, 16 October 2018

Survey spectrum processing

This analysis is continued from here.

Under Display options select 'Stacked Chart View' as seen below. One can see that the peaks in the sandwich emerge and disappear as the depth profile progresses.

Pressing the Automatic Survey ID button will therefore only identify the elements present at each individual level. This is shown below for level 33.

We therefore have to add the remaining peaks manually - this is done by pressing the 'ID' button. In the panel below, select the elements from the periodic table, highlight the peak as shown below with Al2p and press 'Add Peaks'.

Once all the peaks have been added to the peak table, one can create a profile as shown below. Among the profile types choose 'Atomic %'.

In the profile below there is one problem.

As seen above there is only a very small amount of Ti2p. Why is this? Let us zoom in onto the Ti2p peak and see how the fitting goes.