s1= s2 = s3 = s, the resistivity is given by: | s1= s2 = s3 = s, the resistivity is given by: | ||
ρ = 2aπsV/I | ρ = 2aπsV/I Eq.1 | ||
where a is the thickness correction factor for thicknesses t equal to or less than half the probe spacing (t/s < 0.5): | where a is the thickness correction factor for thicknesses t equal to or less than half the probe spacing (t/s < 0.5): | ||