= Modification of showerhead in december 2014 =
= Modification of showerhead in december 2014 =
This page describes why the showerhead has been changed and what process changes have been found.
== Motivation ==


The showerhead that distributes the process gasses inside the plasma source was changed. In the old design, the showerhead had a number of 300 µm diameter holes intended to distribute the gas uniformly inside the plasma source. The passage through the holes had a flow resistance that would cause the gasses coming from the MFC in a Bosch process to get mixed up - in this way working against the separation of the etch and dep cycles.
The showerhead that distributes the process gasses inside the plasma source was changed. In the old design, the showerhead had a number of 300 µm diameter holes intended to distribute the gas uniformly inside the plasma source. The passage through the holes had a flow resistance that would cause the gasses coming from the MFC in a Bosch process to get mixed up - in this way working against the separation of the etch and dep cycles.