Specific Process Knowledge/Characterization/SEM: Scanning Electron Microscopy: Difference between revisions
Line 190: | Line 190: | ||
* Variable at Low vacuum (0.1 mbar-2 mbar) | * Variable at Low vacuum (0.1 mbar-2 mbar) | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
* Conductor: 5 Pa | |||
* Standard: 30 Pa | |||
* Charge-up reduction: 50 Pa | |||
<!--|style="background:WhiteSmoke; color:black"| | <!--|style="background:WhiteSmoke; color:black"| | ||
* High vacuum and Low vacuum--> | * High vacuum and Low vacuum--> |
Revision as of 14:50, 24 January 2017
Feedback to this page: click here
Scanning electron microscopy at Danchip
At Danchip there are five SEMs (Scanning Electron Microscopes) that all cover a wide range of needs both in the cleanroom and outside: From fast in-process verification of different process parameters such as etch rates, step coverages or lift-off quality to ultra high resolution images on any type of sample intended for publication.
- The SEM Supra 1 is located in the basement outside the cleanroom. It is serving two purposes: Serving the users that have samples from outside the cleanroom and serving as training tool; all new SEM users with no/little SEM experience must be trained on this tool and gain basic knowledge (typically 10 hours of usage) here before being qualified for training on the SEMs in the cleanroom.
- The SEM 2 and 3 are located in the cleanroom where they serve as general imaging tools for samples that have been fabricated in the cleanroom. Like SEM Supra 1, they are VP models from Carl Zeiss and will produce excellent images on any sample. The possibility of operating at higher chamber pressures in the VP mode makes imaging of bulk non-conducting samples possible. The SEM Supra 2 is also equipped with an airlock and an EDX detector.
- The SEM Leo s a very reliable and rugged instrument that provides high quality images of most samples. It is exclusively dedicated to the users of the Raith E-beam lithography system so general imaging of user samples is not allowed.
- The SEM Tabletop 1 is a tabletop SEM that is located in the basement outside the cleanroom. It has a limited resolution, but it is fast and easy to use, also for non-conducting samples. Training in the others SEMs is not required to use this SEM.
SEM Supra 1, 2 and 3 SEM Leo s all manufactured by Carl Zeiss and have the same graphical user interface and very identical electron optics. But there are there are small hardware and software differences, thus a training is needed for each SEM you want to use.
The SEM Tabletop 1 is manufactured by Hitachi.
Common challenges in scanning electron microscopy
Comparison of SEM's at Danchip
Equipment | SEM LEO | SEM Supra 1 | SEM Supra 2 | SEM Supra 3 | SEM Tabletop 1 | |
---|---|---|---|---|---|---|
Purpose | Imaging and measurement of |
|
|
|
|
|
Other purpose |
|
|
||||
Instrument Position |
|
|
|
|
| |
Performance | Resolution | The resolution of a SEM is strongly dependent on the type of sample and the skills of the operator. The highest resolution is probably only achieved on special samples | ||||
|
|
|
|
| ||
Instrument specifics | Detectors |
|
|
|
|
|
Stage |
|
|
|
|
| |
Electron source | FEG (Field Emission Gun) source |
| ||||
Operating pressures |
|
|
|
|
| |
Options |
|
|
|
|
||
Substrates | Sample sizes |
|
|
|
|
|
Allowed materials |
|
|
|
|
|
Comparison of the SEM's at CEN
Equipment | SEM Inspect S | SEM FEI Nova 600 NanoSEM | SEM FEI Quanta 200 ESEM FEG | FIB-SEM FEI QUANTA 200 3D | Dual Beam FEI Helios Nanolab 600 |
---|---|---|---|---|---|
Purpose |
|
|
|
|
|
Equipment position | CEN Building 314 | CEN Building 314 | CEN Building 314 | CEN Building 307 Room 111 | CEN Building 314 |
Resolution | The resolution of a SEM is strongly dependent on sample type and the operator. Resolution quoted is using sputtered gold on carbon | ||||
•3.0nm at 30kV (SE) •10nm at 3kV (SE) •4.0nm at 30kV (BSE)
•3.0nm at 30kV (SE) • 4.0nm at 30kV (BSE) • > 12nm at 3kV (SE) |
B |
• 0.8 nm at 30 kV (STEM) • 1.0 nm at 30 kV (SE) • 2.5 nm at 30 kV (BSE) - 3.0 nm at 1 kV (SE)
• 3.0 nm at 1 kV (BD mode + BSE)
•2.5 nm at 30 kV (BSE) •3.0 nm at 3 kV (SE)
•1.4 nm at 30 kV (SE) |
•5nm @30kV
•7nm @ 30kV |
•0.8nm @15kV •0.9nm @1kV
•4.5nm @ 30kV | |
Detectors |
|
|
|
|
|
Stage specifications |
|
|
|
|
|
Options | A | B | C | D | E |
Max sample size | Consult with CEN staff as weight, dimensions, pumping capacity and technique all play a roll in the sample size |