Physimeca

The Physicmeca, placed in the III-V laboratory, cleanroom A-5.

Physimeca is a system for deposition of metals through electron-beam evaporation, intended for III-V processing.


Wafers are loaded into the system through the load lock and transferred into the deposition chamber. Physimeca allows you to deposit metals on samples of different sizes and shapes. It will also take samples of almost any shape as long as they can fit on the sample holders; there are three different sample holders, one for chips (arbitrary size), one for 2" wafers and one for 4" wafers. The sample holder can be tilted during deposition, but is normally kept in a horizontal orientation.

Six different metals are present at each time in the Physimeca. The metal combination is changed if needed. A list of currently allowed metals is included in the table below; contact the ThinFilm team (thinfilm@danchip.dtu.dk) if you have wishes for other metals.


The user manual and contact information can be found in LabManager:

Physimeca in LabManager




Equipment performance and process related parameters Physimeca

Purpose Deposition of metals
  • E-beam evaporation of metals
Performance Film thickness
  • 10 Å - 5000 Å*
Available metals and deposition rates (Å/s)
  • Gold (Au) 10
  • Chromium (Cr) 5
  • Platinum (Pt) 5
  • Titanium (Ti) 5
  • Nickel (Ni) 5
  • Germanium (Ge) 5
  • Palladium (Pd) 5
  • Silver (Ag)
Process parameter range Process Temperature
  • Less then 80 oC
Process pressure
  • 2x10-6 mbar
Substrates Batch size
  • 2 or 4" wafers
  • Or several smaller pieces
Substrate material allowed
  • III-V materials
  • Si
Materials allowed on the substrate
  • Silicon oxide
  • Silicon (oxy)nitride
  • Photoresist
  • PMMA
  • Mylar

* For thicknesses above 2000 Å special permission is required (contact thinfilm@danchip.dtu.dk).

Calibration of Physimeca - for experienced users only