[[File:prealign1.jpg|600px|right]]
[[File:prealign1.jpg|600px|right]]


'''1.''' Carefully position the cassette face down on the optical aligner stage; make sure that the cassette is completely aligned with the stage before clamping the cassette to the stage. The hook of the cassette should turn away from yourself, i.e. towards the wall.
'''1.''' Carefully position the cassette face down on the optical aligner stage; make sure that the cassette is completely aligned with the stage before clamping the cassette to the stage. The hook of the cassette should turn away from yourself, i.e. towards D-3.


'''2.''' Open the Pre-Alignment Microscope System (PAMS) tool and Kappa Image Base from the Desktop. In Kappa Image Base, click ‘Open/Close Control Dialog’ in ‘Camera’. Click on the screen-icon to open camera view and the ‘>>’-icon to open the control dialog. If no contact can be reached to the camera via Kappa Image Base, restart the computer (login: cleanroom, password: Renrum12).
'''2.''' Open the Pre-Alignment Microscope System (PAMS) tool and Kappa Image Base from the Desktop. In Kappa Image Base, click ‘Open/Close Control Dialog’ in ‘Camera’. Click on the screen-icon to open camera view and the ‘>>’-icon to open the control dialog. If no contact can be reached to the camera via Kappa Image Base, restart the computer (login: cleanroom, password: Renrum12).
Line 142: Line 142:




'''5.''' Find the P mark at maximum magnification. Click ‘Get P’. Repeat the procedure with the Q mark. Calculate gain and rotation by clicking 'Calculate' and log the results in the Log-window by clicking 'Log result'. Switch off the yellow light after use.
'''5.''' Find the P mark at maximum magnification. Click ‘Get P’. Repeat the procedure with the Q mark. Calculate gain and rotation by clicking 'Calculate' and log the results in the Log-window by clicking 'Log result'. Switch off the yellow light after use. Make sure that the rotation is < 0.5 degrees and Gain is close to 1. If this is not the case, you might have entered wrong design coordinates for P and Q or found the wrong marks on the chips or wafer.




Line 148: Line 148:




'''7.''' In the jdf-file, the theoretical (L-edit/CleWin) position of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual).
'''7.''' In the jdf-file, the design coordinates of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual). In the sdf-file, enter the material center offset out put from PAMS.


<br clear="all">