'''1.''' Carefully position the cassette face down on the optical aligner stage; make sure that the cassette is completely aligned with the stage before clamping the cassette to the stage. The hook of the cassette should turn away from yourself, i.e. towards | '''1.''' Carefully position the cassette face down on the optical aligner stage; make sure that the cassette is completely aligned with the stage before clamping the cassette to the stage. The hook of the cassette should turn away from yourself, i.e. towards D-3. | ||
'''2.''' Open the Pre-Alignment Microscope System (PAMS) tool and Kappa Image Base from the Desktop. In Kappa Image Base, click ‘Open/Close Control Dialog’ in ‘Camera’. Click on the screen-icon to open camera view and the ‘>>’-icon to open the control dialog. If no contact can be reached to the camera via Kappa Image Base, restart the computer (login: cleanroom, password: Renrum12). | '''2.''' Open the Pre-Alignment Microscope System (PAMS) tool and Kappa Image Base from the Desktop. In Kappa Image Base, click ‘Open/Close Control Dialog’ in ‘Camera’. Click on the screen-icon to open camera view and the ‘>>’-icon to open the control dialog. If no contact can be reached to the camera via Kappa Image Base, restart the computer (login: cleanroom, password: Renrum12). | ||
| Line 142: | Line 142: | ||
'''5.''' Find the P mark at maximum magnification. Click ‘Get P’. Repeat the procedure with the Q mark. Calculate gain and rotation by clicking 'Calculate' and log the results in the Log-window by clicking 'Log result'. Switch off the yellow light after use. | '''5.''' Find the P mark at maximum magnification. Click ‘Get P’. Repeat the procedure with the Q mark. Calculate gain and rotation by clicking 'Calculate' and log the results in the Log-window by clicking 'Log result'. Switch off the yellow light after use. Make sure that the rotation is < 0.5 degrees and Gain is close to 1. If this is not the case, you might have entered wrong design coordinates for P and Q or found the wrong marks on the chips or wafer. | ||
| Line 148: | Line 148: | ||
'''7.''' In the jdf-file, the | '''7.''' In the jdf-file, the design coordinates of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual). In the sdf-file, enter the material center offset out put from PAMS. | ||