Specific Process Knowledge/Bonding: Difference between revisions
No edit summary |
No edit summary |
||
Line 33: | Line 33: | ||
|-valign="top" | |-valign="top" | ||
|'''Substrate size''' | |'''Substrate size''' | ||
|Up to 6" | |Up to 6" (aligning only possible for 4" and 6") | ||
|Up to 6" | |Up to 6" (aligning only possible for 4" and 6") | ||
|Up to 6" | |Up to 6" (aligning only possible for 4" and 6") | ||
|-valign="top" | |-valign="top" | ||
|'''Cleaning''' | |'''Cleaning''' |
Revision as of 13:54, 10 March 2008
Choose bonding method
- Eutectic bonding
- Fusion bonding
- Anodic bonding
Comparing the three bonding methods in the EVG NIL
. | Eutectic bonding | Fusion bonding | Anodic bonding |
---|---|---|---|
General description | For bonding two substrates by use of an interphase that makes an eutecticum. | For bonding two identical materials. | For bonding Si and Glass. |
Bonding temperature | Depending on the eutecticum 310C to 400C. | Depending on defects 50C to 400C. | Depending on the voltage 300C to 500C Standard is 400C. |
Annnealing temperature | No annealing | 1000C in the bond furnace C3 | No annealing |
Materials possible to bond | Bonding of substrates is done by use of the eutectica Au/Si, Au/Sn and Ni/Si | Si/Si, SiO/SiO | Si/Pyrex (glass) |
Substrate size | Up to 6" (aligning only possible for 4" and 6") | Up to 6" (aligning only possible for 4" and 6") | Up to 6" (aligning only possible for 4" and 6") |
Cleaning | Cleaning by N2. | Wet chemical cleaning, IMEC. | Cleaning by N2. |
Choose equipment
- Speedline manual spinner - For spinning of PMMA and Topas
- EVG NIL