Atomika SIMS
The Atomika SIMS analyses the composition of a sample by secondary ion mass spectroscopy. By using either oxygen or cesium ions accelerated by a high tension the surface of the sample is sputtered off as ions. These ions are analysed in a mass spectrometer and one can determine the elemental composition as a function of depth. If compared to signals from reference materials one can quantify the atomic composition - in certain cases down to extremely low concentrations (ppm). Doping levels and impurities may be determined.
Please note that no user will be instructed on the SIMS. Danchip staff will run your samples.
The user manual and technical information and contact information can be found in LabManager:
The Atomika SIMS in Labmanager
An overview of the performance of the SIMS
Purpose | Determination of atomic composition |
|
---|---|---|
Performance | Measurement accuracy depends on |
|
Process parameters | Ion gun parameters |
|
Mass spectrometer parameters |
| |
Sample position |
| |
Sample requirements | Substrate material allowed |
|
Substrate size |
| |
Batch size |
|
Typical application of SIMS
SIMS is the most sensitive technique for elemental composition. It is therefore ideal if you want to check doping profiles or for contaminations.
A typical application would be to check the concentration profile of boron doping in silicon. In that case one would put two samples into the SIMS.
- A reference sample with a known boron profile
- A sample