== Rough estimation of exposure time ==
== Rough estimation of exposure time ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


[[File:currentbeamsize.jpg|400px|right]]
[[File:currentbeamsize.jpg|400px|right]]
Line 73: Line 74:


= Mounting of chips or wafers into cassette =
= Mounting of chips or wafers into cassette =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


Authorized users are allowed to unload a cassette from the robot loader (autoloader) and mount their substrate but '''not''' allowed to load the cassette into the loader after mount.
Authorized users are allowed to unload a cassette from the robot loader (autoloader) and mount their substrate but '''not''' allowed to load the cassette into the loader after mount.
Line 124: Line 126:


= Optical pre-alignment of wafers =
= Optical pre-alignment of wafers =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


[[File:prealign1.jpg|600px|right]]
[[File:prealign1.jpg|600px|right]]
Line 145: Line 148:
7. In the jdf-file, the theoretical (L-edit/CleWin) position of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual).
7. In the jdf-file, the theoretical (L-edit/CleWin) position of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual).


Load and unload of cassettes into/from the e-beam writer


= Load and unload of cassettes into/from autoloader =
= Load and unload of cassettes into/from autoloader =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


[[File:IMG_0186.jpg|600px|right]]
[[File:IMG_0186.jpg|600px|right]]
Line 181: Line 181:


= Transfer of cassettes between autoloader and e-beam writer =
= Transfer of cassettes between autoloader and e-beam writer =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


You can load the cassette into the e-beam writer from the loader control program (Ldr) from EBX Menu. This operation requires the autoloader to be in REMOTE.
You can load the cassette into the e-beam writer from the loader control program (Ldr) from EBX Menu. This operation requires the autoloader to be in REMOTE.
Line 203: Line 204:


= Calibration of condition file =
= Calibration of condition file =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


[[File:ebxmenu.jpg|400px]]
[[File:ebxmenu.jpg|400px]]
Line 303: Line 305:


== Calibration procedure ==
== Calibration procedure ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


=== Load and restore a condition file ===
=== Load and restore a condition file ===


*Load and restore a condition file, click ‘Select condition file...’, choose the file from the list, e.g. ‘0.2nA_ap7’ and click OK. Restore the file by clicking ‘RESTOR/Edit Parameter...’. From the RESTOR window, click ‘Excecute’; the machine will now restore the conditions of the colum, this will take a few minutes. Exit the RESTOR window by clicking ‘Cancel’.
*Load and restore a condition file, click ‘Select condition file...’, choose the file from the list, e.g. ‘0.2nA_ap7’ and click OK. Restore the file by clicking ‘RESTOR/Edit Parameter...’. From the RESTOR window, click ‘Excecute’; the machine will now restore the conditions of the colum, this will take a few minutes. Exit the RESTOR window by clicking ‘Cancel’.


=== Execute CURRNT, INITAE, and INITBE ===   
=== Execute CURRNT, INITAE, and INITBE ===   


*Execute ‘CURRNT’ click ‘CURRNT/Execute’. The program ends by stating the average value of five measurements of the e-beam current; write this average current in the logbook.
*Execute ‘CURRNT’ click ‘CURRNT/Execute’. The program ends by stating the average value of five measurements of the e-beam current; write this average current in the logbook.
Line 318: Line 323:


=== Execute 'daily' ===
=== Execute 'daily' ===


*Execute the pre-defined set of sub-programs called ‘daily’: click 'Commands/Batch...'. Click ‘Condition file...’, choose ‘daily’, and click OK and execute. This set of programs takes around 8-10 minutes to run.
*Execute the pre-defined set of sub-programs called ‘daily’: click 'Commands/Batch...'. Click ‘Condition file...’, choose ‘daily’, and click OK and execute. This set of programs takes around 8-10 minutes to run.
Line 346: Line 352:


== Detailed descriptions of subprograms ==
== Detailed descriptions of subprograms ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


=== SFOCUS ===
=== SFOCUS ===
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>




Line 364: Line 372:


=== PDEFBE ===
=== PDEFBE ===
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


By using the (bottom) BE mark (specified in INITBE), the machine measures the gain and rotation of the main deflector ('''P'''rimary '''DEF'''lector using '''BE''' mark).
By using the (bottom) BE mark (specified in INITBE), the machine measures the gain and rotation of the main deflector ('''P'''rimary '''DEF'''lector using '''BE''' mark).
Line 400: Line 409:


=== DISTMEM and DISTBE ===
=== DISTMEM and DISTBE ===
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


DISTMEM and DISTBE measures the distortion of the beam over the entire writing field, i.e. positional errors in the beam over the entire writing field.
DISTMEM and DISTBE measures the distortion of the beam over the entire writing field, i.e. positional errors in the beam over the entire writing field.
Line 472: Line 482:


=== SUBDEFBE ===
=== SUBDEFBE ===
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


=== DRIFT ===
=== DRIFT ===
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


You can choose any mark as a DRIFT mark. In a standard exposure, a (bottom plane) BE mark is used as DRIFT mark, but you can ask the machine to use one of your global marks as well, i.e. P or Q marks. This requires alignment in semi-automatic mode, however.
You can choose any mark as a DRIFT mark. In a standard exposure, a (bottom plane) BE mark is used as DRIFT mark, but you can ask the machine to use one of your global marks as well, i.e. P or Q marks. This requires alignment in semi-automatic mode, however.
Line 495: Line 507:


=== HEIMAP ===
=== HEIMAP ===
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>




Line 526: Line 539:


= Exposure of mgn-files =
= Exposure of mgn-files =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>




Line 540: Line 554:


= Alignment of wafers or chips =
= Alignment of wafers or chips =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>
Alignment of wafer or chip requires specific training.
Alignment of wafer or chip requires specific training.


Line 547: Line 563:


== AGCRG ==
== AGCRG ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


{| cellpadding="2" style="border: 2px solid darkgray;" align="right"
{| cellpadding="2" style="border: 2px solid darkgray;" align="right"
Line 577: Line 594:


== SETWFR ==
== SETWFR ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


[[File:SETWFR.png|600px]]
[[File:SETWFR.png|600px]]
Line 643: Line 661:


== CHIPAL ==
== CHIPAL ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


[[File:CHIPAL.png|500px]]
[[File:CHIPAL.png|500px]]
Line 668: Line 687:


= Troubleshooting =
= Troubleshooting =
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


'''No EBX Menu is open in any of the desktops:''' Open a console (click on arrow above the text ‘CPU’ on bottom menu bar) and from location (DTU)/ type ‘ebxmenu’.
'''No EBX Menu is open in any of the desktops:''' Open a console (click on arrow above the text ‘CPU’ on bottom menu bar) and from location (DTU)/ type ‘ebxmenu’.
Line 679: Line 699:


== Basic unix operations ==
== Basic unix operations ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


<pre>
<pre>
Line 690: Line 711:


== Calibration problems ==
== Calibration problems ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


'''No current can be detected when executing CURRNT:''' Use another condition file or call one of the contact persons to get help.
'''No current can be detected when executing CURRNT:''' Use another condition file or call one of the contact persons to get help.
Line 730: Line 752:


== Alignment problems ==
== Alignment problems ==
<span style="font-size: 90%; text-align: right;">[[Specific_Process_Knowledge/Lithography/EBeamLithography/JBX9500Manual#top|Go to top of this page]]</span>


'''SETWFR cannot detect my P and Q marks:''' Unload your wafer/chip, remove the resist from your wafer marks and repeat your exposure with manual alignment:
'''SETWFR cannot detect my P and Q marks:''' Unload your wafer/chip, remove the resist from your wafer marks and repeat your exposure with manual alignment: