Specific Process Knowledge/Characterization/Thickness Measurer: Difference between revisions
No edit summary |
|||
Line 2: | Line 2: | ||
==Thickness measurer== | ==Thickness measurer== | ||
[[Image:Thicknessmeasurer2007.jpg|thumb|300x300px|Thickness Measurer. Positioned in cleanroom | [[Image:Thicknessmeasurer2007.jpg|thumb|300x300px|Thickness Measurer. Positioned in cleanroom D-3]] | ||
The purpose is to measure the thickness of wafers, depths of larger grooves or height of larger mesas. | The purpose is to measure the thickness of wafers, depths of larger grooves or height of larger mesas. | ||
Line 12: | Line 12: | ||
'''[http://www.labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=198 Thickness measurer]''' | '''[http://www.labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=198 Thickness measurer]''' | ||
==Quality Control - Recipe Parameters and Limits== | ==Quality Control - Recipe Parameters and Limits== |
Revision as of 11:38, 28 February 2014
Feedback to this page: click here
Thickness measurer
The purpose is to measure the thickness of wafers, depths of larger grooves or height of larger mesas.
During a KOH etch it can be helpful to ensure no over-etching by making a thickness measurement during the etching.
The user manual, technical information and contact information can be found in LabManager:
Quality Control - Recipe Parameters and Limits
Quality Control (QC) for the Thickness measurer |
The measured standard thickness is 0.1 mm. The measured result has to be within ± 0.005 mm. The QC is preformed once a year. |
Purpose |
Thickness measurer |
|
---|---|---|
Performance |
Thickness resolution |
|
Substrates |
Batch size |
|
Substrate materials allowed |
|