Specific Process Knowledge/Characterization/Profiler/Optical Profiler (Sensofar) acceptance test: Difference between revisions
Appearance
| Line 126: | Line 126: | ||
Treshold: 1% | Treshold: 1% | ||
'''Results | |||
It was repeated 10 Time | '''Results''' | ||
It was repeated 10 Time. | |||
The first measurement is shown [[Media:A4_vsi_50xDI_rep_R001.pdf|here]] | The first measurement is shown [[Media:A4_vsi_50xDI_rep_R001.pdf|here]] | ||
'''This table show the depth value for the 10 measurements''' | |||
{| {{table} border="1" | {| {{table} border="1" | ||
| align="center" style="background:#f0f0f0;"|'''''' | | align="center" style="background:#f0f0f0;"|'''''' | ||
| Line 154: | Line 159: | ||
| 10||344.4 | | 10||344.4 | ||
|- | |- | ||
|Average depth||336.58 | |||
|- | |||
|Repeatability||3% | |||
|} | |} | ||