The purpose of the QC process is to check the intensity and uniformity of the exposure light in the Aligner: MA6-2, using a UV optometer. The optometer is placed at 5 fixed positions in the exposure area on a dedicated QC chuck, and the intensity of the light is measured.
The purpose of the QC process is to check the intensity and uniformity of the exposure light in the Aligner: MA6-2, using a UV optometer. The optometer is placed at 5 fixed positions in the exposure area on a dedicated QC chuck, and the intensity of the light is measured.


The 5 intensity measurements are then used to calculate the average intensity as well as the peak uniformity error: <math>PUE[%] = \frac{min-max}{min+max} \sdot 100</math>.
The 5 intensity measurements are then used to calculate the average intensity as well as the peak uniformity error: <math>PUE[%] = \frac{max-min}{max+min} \sdot 100</math>.