==Quality Control (QC)==
==Quality Control (QC)==
The purpose of the QC process is to check the intensity and uniformity of the exposure light in the Aligner: MA6-2, using a UV optometer. The optometer is placed at 5 fixed positions, in the exposure area on a dedicated QC chuck, and the intensity of the light is measured.
The purpose of the QC process is to check the intensity and uniformity of the exposure light in the Aligner: MA6-2, using a UV optometer. The optometer is placed at 5 fixed positions in the exposure area on a dedicated QC chuck, and the intensity of the light is measured.


The 5 intensity measurements are then used to calculate the average intensity as well as the peak uniformity error: <math>PUE[%] = \frac{min-max}{min+max} \sdot 100</math>.
The 5 intensity measurements are then used to calculate the average intensity as well as the peak uniformity error: <math>PUE[%] = \frac{min-max}{min+max} \sdot 100</math>.
Line 142: Line 142:


'''Uniformity accept limit:'''<br>
'''Uniformity accept limit:'''<br>
The uniformity should be adjusted if the uniformity error is greater than 2%.
The uniformity should be adjusted if the peak uniformity error is greater than 2%.


'''Documentation:'''<br>
'''Documentation:'''<br>