Jump to content

Specific Process Knowledge/Lithography/UVExposure/aligner MA6-2: Difference between revisions

Jehem (talk | contribs)
Jehem (talk | contribs)
Line 101: Line 101:
==Alignment==
==Alignment==
'''Top Side Alignment:'''
'''Top Side Alignment:'''
*TSA microscope standard objectives: 5X, and 10X (20X available)
*TSA microscope standard objectives: 5x, and 10x (20x available)
*TSA microscope special objectives: 11.25X offset (for smaller separation)
*TSA microscope special objectives: 11.25x offset for smaller separation


*Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
*Minimum distance between TSA microscope objectives: 33 mm for standard objectives and 8 mm for the special objectives
**Alignment of smaller separations is possible using the "scan microscope" function
**Alignment of even smaller separations is possible using the "scan microscope" function
*Maximum distance between TSA microscope objectives: 160 mm
*Maximum distance between TSA microscope objectives: 160 mm
*TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)
*TSA microscope travel range: X ±25 mm; Y +20/-75 mm (towards the flat)


'''BackSide Alignment:'''
'''BackSide Alignment:'''
*Minimum distance between BSA microscope objectives: 15 mm
*Minimum distance between BSA microscope objectives: 15 mm
*Maximum distance between BSA microscope objectives: 100 mm
*Maximum distance between BSA microscope objectives: 100 mm
*BSA microscope travel range: X +50mm / -16mm; Y +50mm / -20mm (towards flat)
*BSA microscope travel range: X +50/-16 mm; Y +50/-20 mm (towards the flat)
*BSA chuck view ranges:
*BSA chuck view ranges:
**2": X +/- 8-22mm; Y +/- 0-6mm
**50 mm: X ±8-22 mm; Y ±0-6 mm
**4": X +/- 14-46mm; Y +/- 0-10mm
**100 mm: X ±14-46 mm; Y ±0-10 mm
**6": X +/- 14-69mm; Y +/- 0-10mm (OBS: Max. separation of BSA microscopes is 100mm)
**150 mm: X ±14-69 mm; Y ±0-10 mm