Unless otherwise stated, this page is written by DTU Nanolab internal (Feb 2025)), for more info contact mfarin.
New QC tests on Silicon nitride - PECVD4
Due to recent instability in the High-Frequency (HF) generator, the thickness deposited during the QC HF Nitride process has not been very reliable. To better control it, new tests were performed. Since the recipe uses 25W of HF, and the main variation was the reflected power—oscillating between ~5W and 10W—higher power levels were tested. When using 100W, the reflected power was significantly more stable. Below, some of the data is presented, comparing 25W and 100W of HF power, as well as results from different characterization tools (Ellipsometer and FilmTek).
Data with HF generator=25W





Data with HF generator=100W