===Specifications===  
===Specifications===  
- Particles having a size between 79 nm um and 5 um can be detected. The minimum particle size can only be detected on smooth surfaces such as bare silicon wafers or wafers with thermal SiO2.
- Particles having a size between 79 nm um and 5 um can be detected. The minimum particle size can only be detected on smooth surfaces such as bare silicon wafers or wafers with thermal SiO2.<br />
- The detected particles sizes depends on the used gain setting.<br />
- The detected particles sizes depends on the used gain setting.<br />
- Wafer cassettes are available for 4”, 6” and 8” wafers. For the 4” wafer cassette, an adapter has to be mounted on the indexer.<br />
- Wafer cassettes are available for 4”, 6” and 8” wafers. For the 4” wafer cassette, an adapter has to be mounted on the indexer.<br />
- The wafer thickness should be a standard wafer thickness +/- 200 nm. Wafers that thicker than 1 mm need to get an approval from a responsible person.<br />
- The wafer thickness should be a standard wafer thickness +/- 200 nm. Wafers that thicker than 1 mm need to get an approval from a responsible person.<br />
- Area Count software function which helps to detect the larger particles (>11um) is not currently equipped but it is available for purchase.<br />
- Area Count software function which helps to detect the larger particles (>11um) is not currently equipped but it is available for purchase.<br />