Jump to content

Specific Process Knowledge/Lithography/Strip: Difference between revisions

Jehem (talk | contribs)
Jehem (talk | contribs)
Line 222: Line 222:
'''NB: Use dedicated descum asher Plasma Asher 3: Descum for descumming.'''
'''NB: Use dedicated descum asher Plasma Asher 3: Descum for descumming.'''
-->
-->
==Process gas ratio==
[[File:PA_gas_mix_v1.png|320px|thumb|Ashing rate as function of gas mix ratio.|right]]
The ashing rate is related to the gas mix, usually expressed as percentage of nitrogen of the total amount of gas. Test using a single 100 mm wafer in the center of the process chamber shows that 30-80% nitrogen gives the highest ashing rate.
<b>Test parameters:</b><br>
Total gas flow rate: 500 sccm<br>
Chamber pressure: 1.2 mbar<br>
Power: 1000 W<br>
Processing time: 2 minutes<br>
Temperature (average): 43°C
<br clear="all" />
==Process chamber pressure==
==Process gas flow rate==
==Process power==
==Plasma Asher 1==
==Plasma Asher 1==
<span style="color:red">Plasma asher 1 was decommissioned 2024-12-02.</span>
<span style="color:red">Plasma asher 1 was decommissioned 2024-12-02.</span>