Jump to content

Specific Process Knowledge/Characterization/Stress measurement: Difference between revisions

Reet (talk | contribs)
Reet (talk | contribs)
Line 43: Line 43:
#Measure the thickness of the wafer
#Measure the thickness of the wafer
#Deposit the thin film
#Deposit the thin film
#Make a pre-stress measurement. Measure the wafer bow on one of the profilometers ([[Specific Process Knowledge/Characterization/Profiler#Dektak XTA_new stylus profiler|Dektak XTA_new]] or [[Specific Process Knowledge/Characterization/Profiler#Stylus Profiler (Tencor P17)|P17 Stylus profiler]]). Save the measurement. It is a good idea to measure across most of the wafer (at least along 70% of the wafer length) in two directions perpendicular to each other. Using the P17 profiler you can also measure a radially resolved map of the wafer stress with up to 5° resolution.
#Make a pre-stress measurement. Measure the wafer bow on one of the profilometers ([[Specific Process Knowledge/Characterization/Profiler#Dektak XTA_new stylus profiler|Dektak XTA]] or [[Specific Process Knowledge/Characterization/Profiler#Stylus Profiler (Tencor P17)|P17 Stylus profiler]]). Save the measurement. It is a good idea to measure across most of the wafer (at least along 70% of the wafer length) in two directions perpendicular to each other. Using the P17 profiler you can also measure a radially resolved map of the wafer stress with up to 5° resolution.
#Remove the thin film from one side of the wafer. If it is a single side polished wafer, then remove it on the non-polished side.
#Remove the thin film from one side of the wafer. If it is a single side polished wafer, then remove it on the non-polished side.
#Make a post-stress measurement. Measure the wafer bow again:  
#Make a post-stress measurement. Measure the wafer bow again: