Specific Process Knowledge/Etch/DRIE-Pegasus/Pegasus-4/SiO2 Etch/Cr mask: Difference between revisions

From LabAdviser
Jump to navigation Jump to search
No edit summary
No edit summary
Line 79: Line 79:
File:C10026_03__09.jpg
File:C10026_03__09.jpg
File:C10026_03__10.jpg
File:C10026_03__10.jpg
</gallery>
<gallery caption="SiO2 etch with Cr mask on wafer piece on Si carrier 6 min etch, H2:0sccm, C4F8:13sccm He:225sccm, Pressure:3.35mTorr" perrow="6" widths="200px" heights="150px">
File:C10082_01.jpg
File:C10082_03.jpg
File:C10082_05.jpg
File:C10082_07.jpg
File:C10082_09.jpg
File:C10082_11.jpg
</gallery>
</gallery>

Revision as of 15:49, 25 September 2023

Feedback to this page: click here
Unless otherwise stated, all content in this section was done by Berit Herstrøm, DTU Nanolab

Start parameters, variations noted in the gallery headline Recipe name: no 10 with lower platen power
Coil Power [W] 2500
Platen Power [W] 200
Platen temperature [oC] 20
H2 flow [sccm] 25.6
C4F8 flow [sccm] 25.6
He flow [sccm] 448.7
Pressure Fully open APC valve (8-9 mTorr)
Electromagnetic coils (EM) 'outer coil' / 'inner coil' '2 A' / '30 A'
  • 100 nm Cr mask etched in ICP metal with 500nm DUV neg resist and 65 nm barc.