A detailed manual on the Aeris XRD is found on the [https://wiki.iodp.tamu.edu/display/LMUG/Aeris+Advanced+User+Guide wiki of the International Ocean Discovery Project]. It has helpful hints on adjusting the hardware including choices on beam knife height and slit size. It also has some pictures from the XRDMP Creator software. Some info from there:
A detailed manual on the Aeris XRD is found on the [https://wiki.iodp.tamu.edu/display/LMUG/Aeris+Advanced+User+Guide wiki of the International Ocean Discovery Project]. It has helpful hints on adjusting the hardware including choices on beam knife height and slit size. It also has some pictures from the XRDMP Creator software. Some info from there:


...Scanning at angles less than 5° 2θ you will need to lower the knife (anti scatter).  
..."Scanning at angles less than 5° 2θ you will need to lower the knife (anti scatter)."
.... at low angles you also need a narrower slit (which will reduce intensity but also the irradiated area)
 
.... "at low angles you also need a narrower slit (which will reduce intensity but also the irradiated area)."