LabAdviser/314/Microscopy 314-307/TEM/ETEM/OneView: Difference between revisions

From LabAdviser
< LabAdviser‎ | 314‎ | Microscopy 314-307‎ | TEM‎ | ETEM
Jenk (talk | contribs)
mNo edit summary
Jenk (talk | contribs)
mNo edit summary
Line 8: Line 8:




== Headline text ==
== General info ==
General info:<br />
<br />


- CMOS camera with 4096 x 4096px (pixel size: 15µm) --> 61.4 x 61.4mm sensor size; conversion efficiency @300kV: 26.42 ADC counts/Prim e-<br />
- CMOS camera with 4096 x 4096px (pixel size: 15µm) --> 61.4 x 61.4mm sensor size; conversion efficiency @300kV: 26.42 ADC counts/Prim e- (our camera)<br />
- Imaging mode: 25fps for full 4k x 4k resolution (single frame exposure time: 0.04s); 100fps @ 2k x 2k; 200fps @ 1k x 1k; 300fps @ 512 x 512<br />
- Imaging mode: 25fps for full 4k x 4k resolution (single frame exposure time: 0.04s); 100fps @ 2k x 2k; 200fps @ 1k x 1k; 300fps @ 512 x 512<br />
- Diffraction mode:
- Diffraction mode: about 7fps for full 4k x 4k (single frame exposure time: 0.16s)<br />
- drift correction available<br />
- In ''in situ'' mode: look back option available up to 20s<br />
<br />
 
 
== Operation ==
=== image acquisition ===
 





Revision as of 11:00, 21 September 2021

THIS PAGE IS UNDER CONSTRUCTION

Gatan OneView camera and GMS3


Here are a few information on the OneView camera and the control in GMS3:


General info


- CMOS camera with 4096 x 4096px (pixel size: 15µm) --> 61.4 x 61.4mm sensor size; conversion efficiency @300kV: 26.42 ADC counts/Prim e- (our camera)
- Imaging mode: 25fps for full 4k x 4k resolution (single frame exposure time: 0.04s); 100fps @ 2k x 2k; 200fps @ 1k x 1k; 300fps @ 512 x 512
- Diffraction mode: about 7fps for full 4k x 4k (single frame exposure time: 0.16s)
- drift correction available
- In in situ mode: look back option available up to 20s


Operation

image acquisition

Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU"

Run script


2. Type in the high tension the microscope is running in the pop-up window

Select high tension


3. Don't change the conversion factor; just click on "OK"

Don't change conversion factor