Specific Process Knowledge/Etch/DRIE-Pegasus/Pegasus-2: Difference between revisions

From LabAdviser
Jmli (talk | contribs)
Jmli (talk | contribs)
Line 20: Line 20:
}}
}}
{{Template:Peg2configcontent1
{{Template:Peg2configcontent1
|ItemName= aslkdjf;as
|ItemName= Available gasses and gas chemistry
|ItemConfiguration= afjasdflj
|ItemConfiguration=  
|ItemComment=a;dfkjasl
* C<sub>4<\sub>F<sub>8<\sub> (H2 currently fitted but closed) : 0 sccm
* SF6: 50 sccm
* O2: 50 sccm
* Ar: 283 sccm
* N2: 500 sccm
* He: 11 sccm
* CO2: (It is not in the software)
 
 
|ItemComment=OnlySF6 and O2 are used for Si, PR, and Cr etch. The rest is only make-up
}}
}}
|}
|}

Revision as of 15:10, 11 December 2020

Feedback to this page: click here

This system is a research tool and not available to the users

If you want to get access to the tool, then talk to professor Henry Jansen


Current setup and rules on Pegasus 2

Click here to access older configurations.

The current configuration is

valid from November 2020 to today
Item The currently applied modification Comments
Available gasses and gas chemistry * C4<\sub>F8<\sub> (H2 currently fitted but closed) : 0 sccm
  • SF6: 50 sccm
  • O2: 50 sccm
  • Ar: 283 sccm
  • N2: 500 sccm
  • He: 11 sccm
  • CO2: (It is not in the software)
OnlySF6 and O2 are used for Si, PR, and Cr etch. The rest is only make-up

Access to Pegasus 2 configuration templates

Pegasus 2 configuration table version 1