Equipment performance: added QC section
Line 92: Line 92:
*Any standard cleanroom material and samples from the Laser Micromachining tool and the Polymer Injection Molding tool
*Any standard cleanroom material and samples from the Laser Micromachining tool and the Polymer Injection Molding tool
|-  
|-  
|}
===Quality control of length measurement accuracy [[Image:section under construction.jpg|70px]]===
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for DektakXTA'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=2493&mach=304 QC procedure for SEM Supra 1]<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=275 Newest QC data for SEM Supra 1]<br>
{| {{table}}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:300px"
! Detector
! Measurement settings
! Magnification
! Measured dimensions
! Calibration limit
! Action limit
|-
|SE2
| EHT 10 kV, WD 10 mm
| 2.500 k
| 70x70 microns
| ± 2 %
| ± 3 %
|-
|SE2
| EHT 10 kV, WD 10 mm
| 17.000 k
| 10x10 microns
| ± 2 %
| ± 3 %
|-
|SE2
| EHT 5 kV, WD 5 mm
| 17.000 k
| 10x10 microns
| ± 2 %
| ± 3 %
|-
|}
|-
|}
|}
|}