Specific Process Knowledge/Lithography/Aligners/Aligner: Maskless 02 processing: Difference between revisions

Taran (talk | contribs)
Taran (talk | contribs)
Line 24: Line 24:
The correct way to determine the best dose-defocus settings is to generate a so-called Bossung plot (known from projection lithography), which plots the printed linewidth as a function of dose and defocus. From this, the most stable region of parameter space is chosen, i.e. the region where the linewidth changes the least when dose and defocus changes. Any deviation from the design linewidth can be corrected using the CD bias parameter. This typically involves SEM imaging of resist cross-sections, and quickly becomes time consuming. However, in most cases, inspection of a dose-defocus matrix (easily generated using the series exposure function) in an optical microscope will get you most of the way.  
The correct way to determine the best dose-defocus settings is to generate a so-called Bossung plot (known from projection lithography), which plots the printed linewidth as a function of dose and defocus. From this, the most stable region of parameter space is chosen, i.e. the region where the linewidth changes the least when dose and defocus changes. Any deviation from the design linewidth can be corrected using the CD bias parameter. This typically involves SEM imaging of resist cross-sections, and quickly becomes time consuming. However, in most cases, inspection of a dose-defocus matrix (easily generated using the series exposure function) in an optical microscope will get you most of the way.  


'''Data is for Si and optical autofocus unless otherwise stated'''
'''Data represent dose-defocus optimization on Si using optical autofocus unless otherwise stated'''


{|border="1" cellspacing="1" cellpadding="7" style="text-align:left;"  
{|border="1" cellspacing="1" cellpadding="7" style="text-align:left;"