Line 204: Line 204:
'''Result of using only the alignment tool, pushing it to the extremes:'''
'''Result of using only the alignment tool, pushing it to the extremes:'''


-11.08 to 5.54mRad, corresponding to -0.6 to 0.3°, average -0.1±0.5° (6 measurements, including 2 without pushing the alignment tool). Measured using pneumatic autofocus.
-11.08 to 5.54mRad, corresponding to -0.6 to 0.3°, average -0.1±0.5° (6 measurements, including 2 without pushing the alignment tool). Measured using pneumatic autofocus. This is comparable to the ±1° flat-to-crystal alignment spec of the SEMI wafer standard.


It would seem that applying the flat angle measured with optical autofocus to the print introduces more error than relying only on the alignment tool.
It would seem that applying the flat angle measured with optical autofocus to the print introduces more error than relying only on the alignment tool.