==When the deposition is on one side of the wafer==
==When the deposition is on one side of the wafer==
#Make a pre-measurement. Measure the wafer bow on one of the stylus profilometers ([[Specific Process Knowledge/Characterization/Profiler#Dektak XTA_new stylus profiler|Dektak XTA_new]] or [[Specific Process Knowledge/Characterization/Profiler#Dektak _8 stylus _profiler|Dektak 8]]). Save the measurement. It is a good idea to measure across most of the wafer (at least along 70% of the wafer diameter) in two directions perpendicular to each other.
#Make a pre-stress measurement. Measure the wafer bow on one of the stylus profilometers ([[Specific Process Knowledge/Characterization/Profiler#Dektak XTA_new stylus profiler|Dektak XTA_new]] or [[Specific Process Knowledge/Characterization/Profiler#Dektak _8 stylus _profiler|Dektak 8]]). Save the measurement. It is a good idea to measure across most of the wafer (at least along 70% of the wafer diameter) in two directions perpendicular to each other.
#Measure the thickness of the wafer
#Measure the thickness of the wafer
#Deposit the thin film
#Deposit the thin film
#Measure the thickness of the thin film (ex. using the FilmTek or the Ellipsometer).
#Measure the thickness of the thin film (ex. using the FilmTek or the Ellipsometer).
#Make a post-measurement. Measure the wafer bow again:  
#Make a post-stress measurement. Measure the wafer bow again:  
#*On the same profilometer as used for the pre-measurement.  
#*On the same profilometer as used for the pre-measurement.  
#*With the same recipe as used for the pre-mesurement.  
#*With the same recipe as used for the pre-mesurement.  
#*On the same position(s) on the wafer as the pre-measurement .  
#*On the same position(s) on the wafer as the pre-measurement.  
#Use the program for stress measurement in the profilometer software on the profilometer you used for the measurement. Use both the pre-stress measurement and the post-stress measurement.
#Use the program for stress measurement in the profilometer software. Use both the pre-stress measurement and the post-stress measurement.
#*The profiler program asks for substrate elasticity (choose the substrate type for example Si(100)), substrate thickness, film thickness and the name of the pre-stress measurement.
#*The profiler program asks for a substrate elasticity (choose the substrate type, e.g. Si(100)), substrate thickness, film thickness and the name of the pre-stress measurement.


==When the deposition is on both sides of the wafer==
==When the deposition is on both sides of the wafer==