A temperature test of a 100 nm Al deposition at 10 Å/s showed that the back of the wafer stays below 37 <sup>o</sup>C. The same is true for 10 nm Ti/90 nm Au at 10 Å/s. In contrast, deposition of 100 nm Nb at 2 Å/s heated the substrate above 104 <sup>o</sup>C (but less than 110 <sup>o</sup>C). Deposition of 60 nm W at about 1 Å/s heated the substrate to above 110 <sup>o</sup>C. | A temperature test of a 100 nm Al deposition at 10 Å/s showed that the back of the wafer stays below 37 <sup>o</sup>C. The same is true for 10 nm Ti/90 nm Au at 10 Å/s. In contrast, deposition of 100 nm Nb at 2 Å/s heated the substrate above 104 <sup>o</sup>C (but less than 110 <sup>o</sup>C). Deposition of 60 nm W at about 1 Å/s heated the substrate to above 110 <sup>o</sup>C. | ||
You may | You may get some idea of how much the substrate will be heated by looking at the temperature required to give a reasonable vapor pressure for the evaporation. You can find a collection of Honig's vapor pressure curves at the bottom of [https://old.mcallister.com/vacuum.html this page (external link)]. | ||
Please ask us if you need to test the heating of the substrates before making a deposition (contact thinfilm@danchip.dtu.dk). | Please ask us if you need to test the heating of the substrates before making a deposition (contact thinfilm@danchip.dtu.dk). | ||