Here I shortly explain how to calibrate to get work function values:
Here I shortly explain how to calibrate to get work function values:


To get a quantitative value for the work function of a sample material you need to calibrate the AFM tip you are using. This is done by measuring a material with a known work function. We do not have a certified sample for this but we are using a sample that came with the system. it has some lines of Au-Si-Al right next to each other.
To get a quantitative value for the work function of a sample material you need to calibrate the AFM tip you are using. This is done by measuring a material with a known work function. We do not have a certified sample for this but we are using a sample that came with the system. it has some lines of Au-Si-Al right next to each other. The below image is from the Bruker application note: AN10-RevA1-PeakForce_KPFM-appNote.pdf
 
[[File:KPFM workfunction Au Si Al.jpg|400px]]

Revision as of 09:16, 9 January 2018

Feedback to this page: click here


THIS PAGE IS UNDER CONSTRUCTION

KPFM (Kelvin Probe Force Microscopy) measurements can be done with this AFM Icon. It is best for mapping the surface potential on a sample with nanometer resolution but it can also be calibrated to give quantitative values.

Here I shortly explain how to calibrate to get work function values:

To get a quantitative value for the work function of a sample material you need to calibrate the AFM tip you are using. This is done by measuring a material with a known work function. We do not have a certified sample for this but we are using a sample that came with the system. it has some lines of Au-Si-Al right next to each other. The below image is from the Bruker application note: AN10-RevA1-PeakForce_KPFM-appNote.pdf