The IMEC process is used for cleaning of wafers prior to fusion bonding
The IMEC cleaning process is used in circumstances where extraordinary clean wafers are required. It is primarily intended for cleaning wafers prior to fusion bonding. This procedure is based on the IMEC cleaning process: M. Meuris et al. "The IMEC clean: A new concept for particle and metal removal on Si surfaces.", Solid state Technology, Vol. 38, Issue 7, pp 109-113, 1995. Has been slightly modified by Karen Birkelund.
Information about the equipment used in this procedure cand be found in LabManager:
Info page for 'Wafer Cleaning'
Step | Process | Details | Comments | Comments |
---|---|---|---|---|
1 | Pre-bond cleaning of Si wafers prior to fusion bonding. | . |
|
Get CLEAN box for wafers!!! |
2 | Piranha | Mixture: H2SO4:H2O2 (4:1)
Temp: 80oC Time: 5 min (a yellow and a black spot) |
You can use the 'Mask Clean' bath in Wet Bench 04 (in cleanroom D-3). However, this requires planning at least a couple of days in advance, since the bath is primarily used for other purposes!. Alternatively use a very clean glass beaker. The recommended procedure is:
|
Maybe clean the tank the day before! |
3 | Rinse | 2 min. rinse | Rinse you wafers with DI-water. | Put into dedicated wet box for IMEC |
4 | IMEC | Mixture: DI water:5% HF:Isopropanol (100:10:1)
Temp: 25 oC Time: 100 sec (two black spots) |
Make an appointment some days in advance with a Danchip employee to help you with this step!
|
. |
5 | Rinse | 2 min. rinse | Put into dedicated wet box for IMEC | . |
6 | Piranha | Not nitride wafers!
Mixture: H2SO4:H2O2 (4:1) Temp: 80 oC Time: 20 min |
Re-use previous piranha
Makes wafers hydrophilic |
. |
7 | Rinse & spin dry | 5 min rinse | . | . |
8 | Put wafers in new clean carrier box | . | . | . |