Specific Process Knowledge/Etch/DRIE-Pegasus/FAQ/Bonding: Difference between revisions

From LabAdviser
Jump to navigation Jump to search
No edit summary
No edit summary
Line 1: Line 1:
<!--Checked for updates on 3/2-2016 - ok/jmli -->
<!--Checked for updates on 30/7-2018 - ok/jmli -->


'''Feedback to this page:  
'''Feedback to this page:  

Revision as of 13:38, 30 July 2018


Feedback to this page: click here

Question 1
A user tried to etch through a wafer using a carrier and crystalbond as medium (see image below). Even though he ran the standard SPTS process (with 14 % etch load) at 0 degrees, the wafer slid off the carrier. The crystalbond used melts at 65 degrees, is there any type with higher melting point?
The wafer slides off the carrier during process
Answer 1
We usually use Crystalbond 555 or 555HMP, they have flow temps of ~55°C and 65°C respectively. Wafer temp during process can be in the range 77-80°C depending on etch rate, open area etc. but for most cases, the wax reflowing during process is not an issue as it is only a wafer piece which is bonded. Even if the wax reflows, I would expect any major slippage unless the chuck was very much off-level (VERY unlikely) or there was trapped air under the whole wafer which would result in the wafer lifting under vacuum; if the wax then reflows, there could be slippage as the wafer settles back down. I usually perform a check when the wafer is under vacuum in the loadlock; if it looks lifted, I remove it, back to the hotplate to reposition and try to work the trapped air out.

Source: Kerry Roberts, SPTS