From this plot several points are open for discussion:
From this plot several points are open for discussion:
* '''Which peaks to add''': The two spin-orbit doublets Si2p and Cu2p should be treated differently. The Si2p peak is treated as one peak because of the small spin-orbit splitting  
* '''Which peaks to add''': The two spin-orbit doublets Si2p and Cu2p should be treated differently. The Si2p peak is treated as one peak because of the small spin-orbit splitting (the splitting is 0.63 eV less than the 1 eV step size of the survey spectrum, see [[Specific Process Knowledge/Characterization/XPS/Processing/Basics/2peaktable|table in the next section]])


From here the data may be exported to Excel or Word by using the options on the 'W Reporting' tab next to the 'Display Modes' in the upper right corner of the GUI.
From here the data may be exported to Excel or Word by using the options on the 'W Reporting' tab next to the 'Display Modes' in the upper right corner of the GUI.