Specific Process Knowledge/Lithography/EBeamLithography/JBX9500Manual: Difference between revisions
Appearance
| Line 139: | Line 139: | ||
6. Save the alignment information (File/Save). The file should be saved under your name and date, eg ‘mettekjan312012.txt’, in the folder ‘C:\ | 6. Save the alignment information (File/Save). The file should be saved under your name and date, eg ‘mettekjan312012.txt’, in the folder ‘C:\temp’. This file can be transferred to the e-beam computer DCH1352 by email or dropbox. | ||
7. In the jdf-file, the theoretical (L-edit/CleWin) position of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual). | 7. In the jdf-file, the theoretical (L-edit/CleWin) position of the P and Q marks should be defined. In the sdf-file the mark-detection should be set to semi-automatic 'S' (see section 7 in this manual and the separate sdf- and jdf-file preparation manual). | ||
= Load and unload of cassettes into/from autoloader = | = Load and unload of cassettes into/from autoloader = | ||