Line 400: Line 400:




'''If the program does not find the mark''', try a larger scan width and execute AGCRG again, and/or try to scan 'Y only' instead.


|Open and edit AGCRG, in the program window '''1'''Enter the measured (from pre-aligner output) P mark position '''2''' Click 'RG mark detect condition: Setting' and adjust scan width and position. Use a large scan width (200-600 microns). In 'Scan type' choose 'fine scan', 'box scan' and choose the correct mark (cross or L-shaped). Start with an 'X only' scan.
'''3''' Save and execute AGCRG ||If the program finds the mark, ask the program to transfer the gain settings to SETWFR P (both rough and fine) or SETWFR Q. Scan again with same settings but choose 'Y only' in the 'RG mark detect conditions'.
If the program does not find the mark, try a larger scan width and execute AGCRG again, and/or try to scan 'Y only' instead.
If the program does not find the mark you can find it manually using the SEM. Remember to minimise the use of SEM:
If the program still does not find the mark you can find it manually using the SEM. Remember to minimise the use of SEM:


*Open the Stage program from EBX menu
*Open the Stage program from EBX menu