Specific Process Knowledge/Wafer cleaning/7-up & Piranha: Difference between revisions
Line 39: | Line 39: | ||
|Sulfuric acid (98%) and Ammonium sulfate | |Sulfuric acid (98%) and Ammonium sulfate | ||
|Sulfuric acid (98%) and Ammonium sulfate | |Sulfuric acid (98%) and Ammonium sulfate | ||
|Sulfuric acid (98%) and Hydrogen peroxide (30%) in the ratio 4:1 First add H<sub>2</sub>SO<sub>4</sub> into a glass beaker then add H<sub>2</sub>O<sub>2</sub><sup>{{fn|1}}</sup>. | |Sulfuric acid (98%) and Hydrogen peroxide (30%) in the ratio 4:1. First add H<sub>2</sub>SO<sub>4</sub> into a glass beaker then add H<sub>2</sub>O<sub>2</sub><sup>{{fn|1}}</sup>. | ||
|- | |- | ||
|-style="background:WhiteSmoke; color:black" | |-style="background:WhiteSmoke; color:black" |
Revision as of 09:03, 17 March 2014
Feedback to this page: click here
Cleaning of wafers or masks
Cleaning of wafers or masks with sulphuric acid can be done either in a dedicated tank "7-up" or in the fume hood in a beaker "Piranha". Both 7-up and Piranha removes heavy organics. Always use one of these after KOH etch or hot phosphoric acid etch to remove alkali ions before further processing. 7-up and Piranha are also used as cleaning solutions after stripping resist.
The user manual, user APV and contact information can be found in LabManager: 7-up 6" info page in LabManager or 7-up (masks) info page in LabManager
The APV for working with piranha in fumehood can be found by clicking here
Comparing the 7-up cleans and Piranha clean
7-up wafers | 7-up Masks | Piranha | |
---|---|---|---|
General description |
Cleaning of wafers using the dedicated tank in cleanroom D3. |
Cleaning of masks using the dedicated tank in cleanroom D3. |
Cleaning of wafers using a beaker in the fumehood in cleanroom B1. Used for glass wafers or wafers with metal or other materials that you are not allowed to put in the 7-up for wafers or masks. |
Chemical solution | Sulfuric acid (98%) and Ammonium sulfate | Sulfuric acid (98%) and Ammonium sulfate | Sulfuric acid (98%) and Hydrogen peroxide (30%) in the ratio 4:1. First add H2SO4 into a glass beaker then add H2O21. |
Process temperature | 80 oC | 80 oC | ~70 oC the chemicals will heat up to working temperature during mixing, therefore be careful! First add H2SO4 into a glass beaker then add H2O21. |
Process time | 10 min. | 10 min. | 10 min. |
Allowed materials |
|
|
All materials (in beaker). |
Batch size |
1-25 4" or 6" wafers |
1-25 4" or 6" wafers or 1-4 5" masks |
1-5 4" wafer at a time |
Size of substrate |
4-6" wafers |
4-6" wafers or 5" masks |
All sizes that can fit into the beaker in a dedicated holder |
Note 1: In preparing a solution involving an acid, always add the acid last. The exception to this rule is Piranha, in which case you add H2O2, which is a very strong oxidant, to H2SO4, which is a very strong acid. This is done because it is potentially explosive and at the very least will cause the solution to become very warm.