Specific Process Knowledge/Characterization/Stress measurement: Difference between revisions

Knil (talk | contribs)
Knil (talk | contribs)
Line 8: Line 8:
==When the deposition is on one side of the wafer==
==When the deposition is on one side of the wafer==
#Measure the wafer bow on one of the profilometers ([[Specific Process Knowledge/Characterization/Profiler#Dektak XTA_new stylus profiler|Dektak XTA_new]] or [[Specific Process Knowledge/Characterization/Profiler#Dektak _8 stylus _profiler|Dektak 8]]). Save the measurement. It is a good idea to measure across most of the wafer (at least along 70% of the wafer length) in two directions perpendicular to each other.
#Measure the wafer bow on one of the profilometers ([[Specific Process Knowledge/Characterization/Profiler#Dektak XTA_new stylus profiler|Dektak XTA_new]] or [[Specific Process Knowledge/Characterization/Profiler#Dektak _8 stylus _profiler|Dektak 8]]). Save the measurement. It is a good idea to measure across most of the wafer (at least along 70% of the wafer length) in two directions perpendicular to each other.
#Measure the thickness  of the substrate
#Deposite the thin film
#Deposite the thin film
#Measure the thickness of the thin film (ex. using the FilmTek or the Ellipsometer is possible).
#Measure the wafer bow again:  
#Measure the wafer bow again:  
#*On the same profilometer as used for the pre-measurement.  
#*On the same profilometer as used for the pre-measurement.  
#*With the same recipe as used for the pre-mesurement.  
#*With the same recipe as used for the pre-mesurement.  
#*On the same position/positions on the wafer as the pre-measurement .  
#*On the same position/positions on the wafer as the pre-measurement .  
#Measure the thickness  of the substrate
#Measure the thickness of the thin film (ex. using the FilmTek or the Ellipsometer is possible).
#Use the program for stress measurement in the profilometer software on the profilometer you used for the measurements. Use both the pre-stress measurement and the post-stress measurement.
#Use the program for stress measurement in the profilometer software on the profilometer you used for the measurements. Use both the pre-stress measurement and the post-stress measurement.
#*The dektak program asks for substrate elasticity (choose the substrate type for example Si(111)), substrate thickness, film thickness and the name of the pre-stress measurement.


==When the deposition is on both sides of the wafer==
==When the deposition is on both sides of the wafer==