Specific Process Knowledge/Characterization/SIMS: Secondary Ion Mass Spectrometry: Difference between revisions
Line 27: | Line 27: | ||
!style="background:silver; color:black" align="left"|Performance | !style="background:silver; color:black" align="left"|Performance | ||
|style="background:LightGrey; color:black"|Measurement accuracy depends on | |style="background:LightGrey; color:black"|Measurement accuracy depends on | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
* Which atoms to be analysed | * Which atoms to be analysed | ||
* The sample morphology (flat samples are much more suited than particles) | * The sample morphology (flat samples are much more suited than particles) | ||
|- | |- | ||
!style="background:silver; color:black" align="left"|Process parameters | !style="background:silver; color:black" align="left" rowspan="3"|Process parameters | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| Ion gun parameters | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
* | * Acceleration voltage, lens parameters | ||
* | * Gas inlet pressures, apertures, | ||
|- | |- | ||
!style="background:silver; color:black" align="left" rowspan="3" |Sample requirements | !style="background:silver; color:black" align="left" rowspan="3" |Sample requirements |
Revision as of 13:34, 24 January 2014
Feedback to this page: click here
Atomika SIMS
The Atomika SIMS analyses the composition of a sample by secondary ion mass spectroscopy. By using either oxygen or cesium ions accelerated by a high tension the surface of the sample is sputtered off as ions. These ions are analysed in a mass spectrometer and one can determine the elemental composition as a function of depth. If compared to signals from reference materials one can quantify the atomic composition - in certain cases down to extremely low concentrations (ppm). Doping levels and impurities may be determined.
Please note that no user will be instructed on the SIMS. Danchip staff will run your samples.
The user manual(s), quality control procedure(s) and results, user APV(s), technical information and contact information can be found in LabManager:
The Atomika SIMS in Labmanager
An overview of the performance of the SIMS
Purpose | Determination of atomic composition |
|
---|---|---|
Performance | Measurement accuracy depends on |
|
Process parameters | Ion gun parameters |
|
Sample requirements | Substrate material allowed |
|
Substrate size |
| |
Batch size |
|